DRAFT 20 June 2005
Jaundice Meter (JM-103) User Manual (usr070)
Page B - 5
Graph 3 – Regression Analysis, Doctors’ Office #1
JM-103 Regression Analysis
Doctor's office No.1
-5
0
5
10
15
20
25
30
0
5
10
15
20
25
30
Serum Bilirubin TSB (mg/dl)
J
M
-103 R
eadi
n
g
TcB
M
g
/d
l
Regression Line
Upper Limit
Low er Limit
Data Points
Regression Analysis Statistics
Sample Size (n) = 133
.
Y = -0.636 + 0.977 X
Root Mean Square Error (MSE) = 1.572
R-Square = 0.804
95% Confidence Interval (CI) for the Intercept = (-1.526, 0.254)
95% CI for the Slope = (0.894, 1.061)