DRAFT 20 June 2005
Page 2 - 12
Jaundice Meter (Model JM-103) User Manual (usr070)
Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Jaundice Meter (JM-103) is intended for use in the electromag-
netic environment specified below. The customer or user of the unit
should ensure that the unit is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—
Guidance
Electrostatic
discharge
(ESD)—
IEC 61000-
4-2
± 6 kV con-
tact
± 8 kV air
± 6 kV con-
tact
± 8 kV air
The floors should be
wood, concrete, or
ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical
fast tran-
sient/burst
—IEC
61000-4-4
± 2 kV for
power sup-
ply lines
± 2 kV for
power sup-
ply lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge—IEC
61000-4-5
± 1 kV
differential
mode
± 1 kV
differential
mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Содержание JM-103
Страница 2: ......
Страница 24: ...DRAFT 20 June 2005 Page 2 16 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Страница 30: ...DRAFT 18 May 2005 Page 3 6 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Страница 44: ...DRAFT 20 June 2005 Page 5 8 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Страница 48: ...DRAFT 18 May 2005 Page 6 4 Jaundice Meter Model JM 103 User Manual usr070 NOTES...
Страница 68: ...DRAFT 20 June 2005 Page A 18 Jaundice Meter JM 103 User Manual usr070 NOTES...
Страница 90: ...DRAFT 9 Nov 2004 Page C 14 Jaundice Meter JM 103 User Manual usr070 NOTES...
Страница 91: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 14 04...