background image

20

21

102

The  manufacture  of  the  devices  takes  place  according  to  the  terms  of  the 
European standard for blood-pressure measuring devices (see technical data)
You  must  consult  your  specialist  dealer  or  chemist  if  there  are  technical 
problems  with  the  blood-pressure  instrument. 

Never  attempt  to  repair  the 

instrument yourself!

a) Do not expose the device to extreme temperatures, humidity, dust, or direct
     sunlight.
b) The cuff contains a sensitive air-tight bubble. Handle it carefully and avoid all
     types of straining through twisting or buckling.
c) Clean the device with a soft and a dry cloth. Do not use petrol, thinners, or similar
     other solvents. Spots on the cuff can be removed carefully with damp cloth and
     soapsuds. The cuff must not be washed!
d) Do not drop the instrument or treat it roughly in any way.
e)  Never open the device! Otherwise, the manufacturer calibration becomes
     invalid!

Periodical recalibration

Sensitive  measuring  devices  must  be  checked  for  accuracy  time  to  time.  We 
therefore recommend a periodical inspection of the static pressure display every 2 
years.  Your  specialist  dealer  would  be  pleased  to  provide  more  extensive 
information about this.

7. CARE AND MAINTENANCE, RE-CALIBRATION

Safety  and  protection

·

This instrument may be used only for the purpose described in this booklet. The 
manufacturer cannot be held responsible for the damage caused by incorrect 
application.

·

This instrument comprises sensitive components therefore must be treated 
with caution. Observe the storage and operating conditions described in the 
“Technical specifications” section!

·

Protect  it  from  water  and  moisture,  extreme  temperatures,  impact  and 
dropping, contamination and dust, direct sunlight, heat and cold, etc.

·

The cuff is sensitive and must be handled with care.

·

Only pump up the cuff once fitted.

·

Do  not  use  the  instrument  close  to  strong  electromagnetic  fields  such  as 
cellphones or radio installations.

·

Do not use the instrument if you think it is damaged or notice anything unusual.

·

If the instrument is not going to be used for a prolonged period, the batteries 
should be removed. Read the additional safety instructions in the individual 
sections of this booklet. 

·

Ensure that children do not use the instrument unsupervised: some parts are 
small enough to be swallowed.

·

Must use the recognized accessories, detachable parts and materials, if the 
use of other parts or materials can degrade minimum safety.

·

A warning to remove primary batteries if the instruments is not likely to be used 
for some time.

8. SAFETY, CARE AND DISPOSAL

Instrument care

Clean the instrument only with a soft and dry cloth.

Disposal

        Batteries and electronic instruments must be disposed off in accordance with
        the locally applicable regulations, not with domestics waste.

Содержание AFIB Pro

Страница 1: ...ynamically and tirelessly to provide the best of the medical devices to our clients The products that we have to offer are suitably designedforuseathomes laboratoriesandhospitals Our ground breaking solutions allow you to monitor your health in the easiest ways possible In today s era when all of our lives are too hassled to handle it becomes a bit difficult to pay attention to our health But it h...

Страница 2: ...EASUREMENT 4 1 Before the measurement 4 2 Common sources of error 4 3 Fitting thecuff 4 4 Measuring procedure 4 5 Discontinuing a measurement 4 6 Memory storageandrecallofthemeasurements 4 7 Memory cancellationofallmeasurementsattention 5 APPEARANCE OF THE HEART ARRHYTHMIA INDICATOR FOR EARLY DETECTION 6 ERROR MESSAGES MALFUNCTIONS 4 7 9 11 10 17 18 10 3 7 CARE AND MAINTENANCE RECALIBRATION 8 SAFE...

Страница 3: ... a safe place For further questions on the subject of blood pressure and its measurement pleasecontactyourdoctor a TheLCDScreenofAFIBPro 102 The LCD screen displays the systolic and diastolic blood pressure measurements along with AFIB and heart rate It also displays previously recorded measurements and the date and time when the appropriate button ispressed 1 3 ImportantInformationaboutSelf Measu...

Страница 4: ...gular self check with your blood pressure monitor is recommended In this way you can detect possible changes in your values early and react appropriately If you are undergoing medical treatment to control your blood pressure please keep a record of the level of your blood pressure by carrying out regular self measurements at specific times of the day Show these values to your doctor Never use the ...

Страница 5: ...ng periods please remove thebatteriesfromthedevice 102 Table for classifying blood pressure values unit mmHg according to World HealthOrganization Range broadcasting Systolic Blood pressure DiastolicBlood pressure Measures Optimal Self check Normal Slight hypertension between 120 and129 between 140 and159 between100 and120 between 80 and84 between 90 and99 between 60 and80 Self check Seek medical ...

Страница 6: ...Settingdate time Howtoenterintothesettingmode Ÿ After batteries are inserted the device will enter the setting mode automatically Ÿ Press and hold ON OFF button for 5 seconds the setting mode will appear Thesettingprocedureasfollows It is essential to set date and time Otherwise you will not be able to save your measuredvaluecorrectlywithadateandtime Use buttonasadjustmentbutton Ifyoupressandhold ...

Страница 7: ...per to avoidairleakage a The distance between the edge of cuff andtheelbowshouldbeapprox 2 3cm b Secure the cuff with the Velcro fastener so that it lies Comfortably and not too tight whereby no space should remain betweenthecuffandthearm Note Comparable blood pressure measurements require the same conditions Theseare alwaysquietconditions c Lay the arm on a table with the palm upwards Support the...

Страница 8: ... 128 Diastole 70 Pulse 80 arrhythmia detected Example2 The measured systolic and diastolic blood pressure valuesaswellasthepulsearenowdisplayed Example Fig Systole128 Diastole70 Pulse80 Afibdetected In Standard Single Mode a To slide the AF switch on the side of the device downwards to position 1 Press the ON OFF button the pump begins to inflate the cuff In the display the increasing cuff pressur...

Страница 9: ...s the display will show the symbol CL and then release the button To permanently clear the memory press the MEMORY button while CL is flashing Thissymbol indicatesdetectionofpulseirregularitiesduringthemeasurement In this case the result may deviate from your normal blood pressure In most cases this is no cause for concern However if the symbol appears on a regular basis e g severaltimesaweekwithm...

Страница 10: ...ment produces a different value although the instrument functions normally and thevaluesdisplayedarenormal Blood pressure measured differs from those values measured by the doctor 1 Checkthepositioningofthecuff 2 Measure the blood pressure again in peace and quiet under observance of the detailsmadeunderpoint5 1 Please read the following information and the points listed under Common s o u r c e s...

Страница 11: ...se described in this booklet The manufacturer cannot be held responsible for the damage caused by incorrect application This instrument comprises sensitive components therefore must be treated with caution Observe the storage and operating conditions described in the Technicalspecifications section Protect it from water and moisture extreme temperatures impact and dropping contaminationanddust dir...

Страница 12: ...ss IIa havebeenfulfilled MeasurementProcedure Oscillometric corresponding to Korotkoff method Phase I systolic PhaseV diastolic Display Digitaldisplay Measuring range Pressure 30 to 280 mmHg in 1 mmHg increment Pulse 40 to 199 beat minute Measuringresolution 1mmHg Inflation Automaticinflationbyinternalpump Memoryfunction 2x120memoriesfor2users SYS DIA Pulse Decompression Constantexhaustvalvesystem...

Страница 13: ...ntact 8 kV air Immunity test IEC60601 1 2 test level Compliance level Electromagnetic environment guidance Electromagnetic Immunity IEC60601 1 2 Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Emission Test Compliance Electromagnetic Environment RF emission CISPR 11 Dr Trust AFIB Pro Blood Pressure Monitor 102...

Страница 14: ...8 3 A m Not applicable Not applicable Note UT is the a c mains voltage prior to application of the test level Immunity test IEC60601 1 2 test level IEC60601 1 2 test level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 80 AM 2Hz 3 Vrms P o r t a b l e a n d m o b i l e R F communications equipment should be used no closer to any part of Dr Trust AFIB Pro B...

Страница 15: ...ch as reorienting or relocating Dr Trust AFIB Pro Blood PressureMonitor102 b Overthefrequencyrange150kHzto80MHz fieldstrengthsshouldbelessthan3 V m Recommended Separation Distances Recommended separation distance between portable and mobile RF communicationsequipmentandDrTrustAFIBProBloodPressureMonitor102 Dr Trust AFIB Pro Blood Pressure Monitor 102 is intended for use in an electromagnetic envir...

Страница 16: ...Video www drtrustusa com 102 CONTACT ADDRESS USA Nureca INC USA 276 5th Avenue Suite 704 397 New York NY 10001 USA INDIA Corporate Office Mumbai Nureca Limited 128 Gala Number Udyog Bhavan 1st Floor Sonawala Lane Goregaon East Mumbai City Maharashtra 400063 Contact us India 91 7527013265 91 9356658436 Website www drtrust in Corp Website www nureca com Customer Care Executive Email customercare nur...

Отзывы: