In AF Mode
In AF mode, 2 measurements are automatically taken
in succession and the result is then automatically
analyzed and displayed. Because blood pressure
constantly fluctuates, a result determined in this way
is more reliable than one produced by a single
measurement.
·
After pressing the ON/OFF button, the
symbol appears in the display.
·
The bottom, left hand section of the display shows
numbers 1, 2 to indicate which of the 2
measurements is currently being taken.
·
There is a break of 15 seconds between the
measurements. A countdown indicates the
remaining time.
·
The individual results are not displayed. Your
blood pressure will only be displayed after the 2
measurements are taken.
·
Do not remove the cuff between measurements.
·
If one of the individual measurements was
questionable, a third one is automatically taken.
Measured result
Example 1.
The measured systolic and diastolic blood-pressure
values as well as the pulse are now displayed.
Example (Fig.):
Systole 128, Diastole 70, Pulse 80, arrhythmia
detected.
Example 2.
The measured systolic and diastolic blood-pressure
values as well as the pulse are now displayed.
Example (Fig.):
Systole 128, Diastole 70, Pulse 80, Afib detected.
In Standard Single Mode
a) To slide the AF switch on the side of the device
downwards to position «1». Press the ON/OFF
button, the pump begins to inflate the cuff. In the
display, the increasing cuff-pressure is continually
displayed.
b) After reaching the inflation pressure, the pump
stops and the pressure slowly falls away. The cuff-
pressure is displayed during the measurement.
When the device has detected the pulse, the heart
symbol in the display begins to blink.
c) The measured systolic and diastolic blood-
pressure values as well as the pulse are now
displayed.
In the measuring
After reaching the inflation pressure, the pump stops and the pressure slowly
falls away. The cuff-pressure is displayed during the measurement. When the
device has detected the pulse, the heart symbol in the display begins to blink.
The measurement results are displayed until you switch the device off. If no
button is pressed for 3 minutes, the device switches off automatically.