Measuring range

Pressure: 30 to 280 mmHg (in 1 mmHg 
increment)
Pulse: 40 to 199 beat/minute

Measuring resolution 

1mmHg

Inflation

Automatic inflation by internal pump

Memory function

2 x 120 memories for 2 users (SYS, DIA, Pulse)

Decompression

Constant exhaust valve system

Power Source

4- size “AA” alkaline Batteries

Operation Temperature

5~40°C/41~104°F

Operation Humidity

15%~85%RH maximum

Storage Temperature

-10~55°C/14~131°F

Storage Humidity

10%~95%RH maximum

Dimensions

135×112×71±1.0 mm 

Static accuracy

Pressure: ±3mmHg / Pulse: ±5% of reading

102

If an error occurs during a measurement, the measurement 
is discontinued and a corresponding error code is displayed  
(Example: Error No. 2).

Other possible malfunctions and their elimination

If problems occur when using the device, the following points should be checked 
and if necessary, the corresponding measures are to be taken:

6. ERROR  MESSAGES/MALFUNCTIONS

Error No.

Malfunction

Possible cause(s)

Remedy

ERR 1

ERR 2

ERR 3

No pulse has been detected.
Unnatural pressure impulses influence the measurement

result. Reason: The arm was moved during the Measurement

(Artefact).

The inflation of the cuff takes too long. The cuff is not

correctly seated.

ERR 5

The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take other reading
following directions carefully. Contact you doctor if you
continue to get unusual readings.

Err8

Pressure in cuff is over 290 mmHg.

The  display  remains  empty  when 
the  instrument  is  switched  on 
although the batteries are in place.

1. Check batteries for correct polarity and 
if necessary insert correctly.
2.  If  the  display  is  unusual,  re-insert 
batteries or exchange them.

The  device  frequently  fails  to 
measure the blood pressure values, 
or the values measured are too low 
(too high).

Every  measurement  produces  a 
d i f fe re nt   va l u e   a l t h o u g h   t h e 
instrument  functions  normally  and 
the values displayed are normal.

Blood  pressure  measured  differs 
from those values measured by the 
doctor.

1. Check the positioning of the cuff. 
2.  Measure  the  blood-pressure  again  in 
peace and quiet under observance of the 
details made under point 5.

1.  Please  read  the  following  information 
and  the  points  listed  under  «Common 
s o u r c e s   o f   e r r o r » .   R e p e a t   t h e 
measurement.

Please  note:  Blood  pressure  fluctuates 
continually so successive measurements 
will show some variability.

1.  Record  the  daily  development  of  the 
values  and  consult  your  doctor. 

Please 

note:  Individuals  visiting  their  doctor 
frequently experience anxiety which can 
result  in  a  higher  reading  at  the  doctor 
than  obtained  at  home  under  resting 
conditions.

Further Information

The level of blood-pressure is subject to fluctuations even with healthy people. 
Important thereby is, that comparable measurements always require the same 
conditions (Quiet conditions)

!

If, in spite of observing all these factors, the fluctuations are larger than 15mmHg, 
and/or you hear irregular pulse tones on several occasions, please consult your 
doctor.
For licensing, the device has been subjected to strict clinical tests, by which the 
computer program used to measure the blood-pressure values was tested by 
experienced specialist doctors in Germany. The same computer program is used 
in every individual device, and has thus also been clinically tested.

Содержание AFIB Pro

Страница 1: ...ynamically and tirelessly to provide the best of the medical devices to our clients The products that we have to offer are suitably designedforuseathomes laboratoriesandhospitals Our ground breaking solutions allow you to monitor your health in the easiest ways possible In today s era when all of our lives are too hassled to handle it becomes a bit difficult to pay attention to our health But it h...

Страница 2: ...EASUREMENT 4 1 Before the measurement 4 2 Common sources of error 4 3 Fitting thecuff 4 4 Measuring procedure 4 5 Discontinuing a measurement 4 6 Memory storageandrecallofthemeasurements 4 7 Memory cancellationofallmeasurementsattention 5 APPEARANCE OF THE HEART ARRHYTHMIA INDICATOR FOR EARLY DETECTION 6 ERROR MESSAGES MALFUNCTIONS 4 7 9 11 10 17 18 10 3 7 CARE AND MAINTENANCE RECALIBRATION 8 SAFE...

Страница 3: ... a safe place For further questions on the subject of blood pressure and its measurement pleasecontactyourdoctor a TheLCDScreenofAFIBPro 102 The LCD screen displays the systolic and diastolic blood pressure measurements along with AFIB and heart rate It also displays previously recorded measurements and the date and time when the appropriate button ispressed 1 3 ImportantInformationaboutSelf Measu...

Страница 4: ...gular self check with your blood pressure monitor is recommended In this way you can detect possible changes in your values early and react appropriately If you are undergoing medical treatment to control your blood pressure please keep a record of the level of your blood pressure by carrying out regular self measurements at specific times of the day Show these values to your doctor Never use the ...

Страница 5: ...ng periods please remove thebatteriesfromthedevice 102 Table for classifying blood pressure values unit mmHg according to World HealthOrganization Range broadcasting Systolic Blood pressure DiastolicBlood pressure Measures Optimal Self check Normal Slight hypertension between 120 and129 between 140 and159 between100 and120 between 80 and84 between 90 and99 between 60 and80 Self check Seek medical ...

Страница 6: ...Settingdate time Howtoenterintothesettingmode Ÿ After batteries are inserted the device will enter the setting mode automatically Ÿ Press and hold ON OFF button for 5 seconds the setting mode will appear Thesettingprocedureasfollows It is essential to set date and time Otherwise you will not be able to save your measuredvaluecorrectlywithadateandtime Use buttonasadjustmentbutton Ifyoupressandhold ...

Страница 7: ...per to avoidairleakage a The distance between the edge of cuff andtheelbowshouldbeapprox 2 3cm b Secure the cuff with the Velcro fastener so that it lies Comfortably and not too tight whereby no space should remain betweenthecuffandthearm Note Comparable blood pressure measurements require the same conditions Theseare alwaysquietconditions c Lay the arm on a table with the palm upwards Support the...

Страница 8: ... 128 Diastole 70 Pulse 80 arrhythmia detected Example2 The measured systolic and diastolic blood pressure valuesaswellasthepulsearenowdisplayed Example Fig Systole128 Diastole70 Pulse80 Afibdetected In Standard Single Mode a To slide the AF switch on the side of the device downwards to position 1 Press the ON OFF button the pump begins to inflate the cuff In the display the increasing cuff pressur...

Страница 9: ...s the display will show the symbol CL and then release the button To permanently clear the memory press the MEMORY button while CL is flashing Thissymbol indicatesdetectionofpulseirregularitiesduringthemeasurement In this case the result may deviate from your normal blood pressure In most cases this is no cause for concern However if the symbol appears on a regular basis e g severaltimesaweekwithm...

Страница 10: ...ment produces a different value although the instrument functions normally and thevaluesdisplayedarenormal Blood pressure measured differs from those values measured by the doctor 1 Checkthepositioningofthecuff 2 Measure the blood pressure again in peace and quiet under observance of the detailsmadeunderpoint5 1 Please read the following information and the points listed under Common s o u r c e s...

Страница 11: ...se described in this booklet The manufacturer cannot be held responsible for the damage caused by incorrect application This instrument comprises sensitive components therefore must be treated with caution Observe the storage and operating conditions described in the Technicalspecifications section Protect it from water and moisture extreme temperatures impact and dropping contaminationanddust dir...

Страница 12: ...ss IIa havebeenfulfilled MeasurementProcedure Oscillometric corresponding to Korotkoff method Phase I systolic PhaseV diastolic Display Digitaldisplay Measuring range Pressure 30 to 280 mmHg in 1 mmHg increment Pulse 40 to 199 beat minute Measuringresolution 1mmHg Inflation Automaticinflationbyinternalpump Memoryfunction 2x120memoriesfor2users SYS DIA Pulse Decompression Constantexhaustvalvesystem...

Страница 13: ...ntact 8 kV air Immunity test IEC60601 1 2 test level Compliance level Electromagnetic environment guidance Electromagnetic Immunity IEC60601 1 2 Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Emission Test Compliance Electromagnetic Environment RF emission CISPR 11 Dr Trust AFIB Pro Blood Pressure Monitor 102...

Страница 14: ...8 3 A m Not applicable Not applicable Note UT is the a c mains voltage prior to application of the test level Immunity test IEC60601 1 2 test level IEC60601 1 2 test level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 80 AM 2Hz 3 Vrms P o r t a b l e a n d m o b i l e R F communications equipment should be used no closer to any part of Dr Trust AFIB Pro B...

Страница 15: ...ch as reorienting or relocating Dr Trust AFIB Pro Blood PressureMonitor102 b Overthefrequencyrange150kHzto80MHz fieldstrengthsshouldbelessthan3 V m Recommended Separation Distances Recommended separation distance between portable and mobile RF communicationsequipmentandDrTrustAFIBProBloodPressureMonitor102 Dr Trust AFIB Pro Blood Pressure Monitor 102 is intended for use in an electromagnetic envir...

Страница 16: ...Video www drtrustusa com 102 CONTACT ADDRESS USA Nureca INC USA 276 5th Avenue Suite 704 397 New York NY 10001 USA INDIA Corporate Office Mumbai Nureca Limited 128 Gala Number Udyog Bhavan 1st Floor Sonawala Lane Goregaon East Mumbai City Maharashtra 400063 Contact us India 91 7527013265 91 9356658436 Website www drtrust in Corp Website www nureca com Customer Care Executive Email customercare nur...

Отзывы: