Ceiling and wall attachment
Dr. Mach
Lamps and Engineering
Connecting diagram
6. Mounting the articulated arms
6.1 Ceiling mounting
6.1.1. Mounting the extension arm
Attention!
Disconnect the on-site power supply and protect
it from being switched on again.
•
Remove the fastening devices (retaining ring
1
,
tab washer
2
and disk
∅
38mm
3
) from the sus-
pension tube (tool: mounting plier).
•
Remove the lateral covering plates of the bracket
as described in chapter 6.1.2.
•
Carefully remove the plug
5
from the extension
arm
6
and slide the extension arm onto the single
part extension arm bearing
4
.
Attention!
Danger of injury: If the tab washer is missing, the
retaining ring will loosen and the device may fall out
of the mounting and cause serious injury.
Always install the tab washer.
•
Install the disk
∅
38mm
3
, then insert the tab
washer
2
into the hole
7
and install the circlip
1
.
•
Check the extension arm for firm seating.
6.1.2 Removing the lateral covering
plates
4
7
6
3
2
1
5
59500001
Edition 06
11.03.2003 / Bak Page 25/48