Lamps and Engineering
59080001
Edition 08
06.02.2010 / Bak Page 10/19
4.3.5 Conversion to sterilisable handle
Extent of supply:
−
sterilisable handle sleeve
−
couple
−
one screw M4x35 DIN 912:A2 with safety lake
•
Turn off the lamp.
•
Remove covering cap 18 with a screw driver.
•
Loosen the screw S M4x6 DIN 7985 with a
cross screw driver.
•
Remove screw S M4x6 DIN 7985 and lock
washer ZS
∅
4,3 DIN 6797.
•
Hold the handle, while removing the screw and
the lock washer, to ensure no damaging in the
lamp housing.
•
Remove handle 17 by using a plastic hammer
(see arrow in figure).
•
Put couple 49 on the handle bar 16.
•
Fix couple 49 with the screw S1 M4x35 DIN
912:A2 with safety lake.
•
Push the sterilizable handle sleeve on the cou-
ple 49, so that it is secured by push-button DK.
16
ZS
S
17
18
16
DK
49
S1
Plastic hammer