MACH 120F/120
Ceiling, Wall, Stand model
Dr. Mach
Lamps and Engineering
59080001
Edition 08
06.02.2010 / Bak Page 17/19
9. Electromagnetic compatibility
Table 201 – Guidance and manufacturer´s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS (see
6.8.3.201 a) 3))
1
Guidance and manufacturer´s declaration – electromagnetic emission
2
The MACH 120 / 120 F is intended for use in the electromagnetic environment specified below. The
customer or the user of the MACH 120 / 120 F should assure that it is used in such an environment.
3
Emissions test
Compliance
Electromagnetic environment - guidance
7
Harmonic emissions
IEC 61000-3-2
Class C
8
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
The MACH 120 / 120 F is suitable for use in all
establishments, including domestic establishments and those directly con-
nected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
12
RF emissions
CISPR 15
Complies
The MACH 120 / 120 F is not suitable for interconnection
with other equipment.
Table 202 – Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS (see
6.8.3.201 a) 6))
Guidance and manufacturer´s declaration – electromagnetic immunity
The MACH 120 / 120 F is intended for use in the electromagnetic environment specified below. The customer
or the user of the MACH 120 / 120 F should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output lines
± 2 kV for power
supply lines
not applicable
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(>95 % dip in U
T
)
for 5 sec
< 5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the
MACH 120 / 120 F requires
continued operation during power
mains interruptions, it is recommended
that the MACH 120 / 120 F be powered from
an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE U
T
is the a. c. mains voltage prior to application of the test level.