Quintaflex R96
Ceiling, Wall Model
Dr. Mach
Lamps and Engineering
59250001
Edition 01
21.02.2002 / Bak Page 5/18
2. Operating the lamp Quintaflex R96
2.1 Checking the lamp Quintaflex R96 with light intensity control
The lamp Quintaflex is optionally equipped with the function of electronic light intensity control. You can
adjust the light intensity at the key pad on the cardan bow.
♦
Voltage measurement and setting
Before using the OT-lamps, a voltage measurement and, if necessary, a voltage setting has to be done.
Only in this way a perfect functioning can be ensured.
If the location of the lamp is changed, the voltage measurement and setting must be repeated.
♦
Voltage measurement at the halogen bulbs
To measure the voltage proceed as follows:
Mount the lamp to the ceiling or wall attachment.
Switch on the lamp.
In order to make the measurement, remove the
cover.
The cover is fixed with two screws, as shown in the
figure. Loosen the two screws, remove them and
take off the cover.
The measuring points for checking the voltage are in
the same position at all Quintaflex-lamps.
The measuring points are shown in the figure.
At measuring points
1 + 2
measure the voltage
applied at the halogen bulbs.
The lamp is set to the
maximum light intensity
.
The AC or DC voltage depends on the external power
supply.
Rated voltage 22,8 – 23,8V.
♦
Voltage setting at the halogen bulbs
If the voltage measured is too high or too low, the
value must be adjusted at the power supply.
The setting at the transformer is made by
reconnecting the wires on the secondary side.
Measuring points
1+2
Measuring points
Measuring points
Screw
Screw