background image

 

Quintaflex R96 

Ceiling, Wall Model

 

Dr. Mach

 

Lamps and Engineering

 

 

59250001 

Edition 01 

21.02.2002 / Bak  Page 2/18

 

Internet: 

www.dr-mach.com

, E-mail: 

[email protected]

 

Содержание 5502 3031 27

Страница 1: ... 01 21 02 2002 Bak Page 1 18 DIRECTIONS FOR USE Quintaflex R96 Single Ceiling Lamp_______________________________Order No 5502 3031 27 Order No 5502 3031 30 Ceiling Lamps Combinations Dr Mach GmbH u Co Flossmannstrasse 28 D 85560 Ebersberg Tel 49 0 8092 2093 0 Fax 49 0 8092 2093 50 ...

Страница 2: ...Quintaflex R96 Ceiling Wall Model Dr Mach Lamps and Engineering 59250001 Edition 01 21 02 2002 Bak Page 2 18 Internet www dr mach com E mail info dr mach de ...

Страница 3: ...erilizable handle Page 7 3 2 Lamp head splinter protection disk Page 9 4 Maintenance Page 10 4 1 Adjustments at the ceiling attachment Page 10 4 2 Adjustments at the lamp head Page 12 4 3 Changing of spare parts Page 13 4 3 1 Changing the halogen bulbs Page 13 4 3 2 Changing the fuses Page 13 4 3 3 Changing the filter disk Page 14 4 3 4 Changing the splinter protection disk Page 14 5 Data Page 15 ...

Страница 4: ...nts based on 50 years of experience in the production of operating and doctor s lights Please read the safety instructions and product description carefully before working with these lights for the first time Please follow the separate mounting instructions for ceiling and wall mounted lamps and for fitting the stand foot and pipe 1 Safety instructions Please pay attention to the directions for us...

Страница 5: ...measure the voltage proceed as follows Mount the lamp to the ceiling or wall attachment Switch on the lamp In order to make the measurement remove the cover The cover is fixed with two screws as shown in the figure Loosen the two screws remove them and take off the cover The measuring points for checking the voltage are in the same position at all Quintaflex lamps The measuring points are shown in...

Страница 6: ...Quintaflex R96 Ceiling Wall Model Dr Mach Lamps and Engineering 59250001 Edition 01 21 02 2002 Bak Page 6 18 ...

Страница 7: ...t it as described below Voltage setting at the control board Switch the lamp off The position of the control board is shown in the figure Adjust the standard setting by changing the jumpers see arrow Three settings enable you to ideally set your lamp system J4 and J5 Take off one jumper and put it into position as shown in figure II or III Switch on the lamp Set the lamp to the lowest light intens...

Страница 8: ... low room heights the lamp is delivered with a central spring arm The spring arm has only one joint The coupling for the bow of the lamp body is fitted with a needle bearing The inner ring 4 of this needle bearing has already been pushed on the axis of the lamp bow Remark The instructions mentioned above refer to ceiling and wall models 2 4 Light field adjustment merging of light fields The lamps ...

Страница 9: ...ly sufficiently validated equipment and product specific processes are used for cleaning disinfection and that the validated parameters are complied with in every cycle In addition the hospital clinic hygiene regulations must be observed Cleaning disinfection Cleaning and disinfection must be carried out immediately after use A mechanised process disinfector should be used for cleaning disinfectio...

Страница 10: ...ed handles may be sterilised The handles are placed in a suitable sterilisation pack one way sterilisation pack e g foil paper sterilisation bags single or double pack in accordance with DIN EN 868 ISO 11607 for steam sterilisation and then sterilised Use only the sterilisation procedure listed below for sterilisation Other sterilisation procedures e g ethylene oxide formaldehyde and low temperatu...

Страница 11: ... which can be cleaned with conventional cleaning agents The splinter protection disks 3 are made of a high quality plastic Pay attention to the following during cleaning Wipe the splinter protection disks 3 always with a wet cloth Do not wipe dry Only use disinfectant with less than 20 alcohol After cleaning wipe the splinter protection disks 3 with an antistatic non fluffy cloth Alc 20 2 3 ...

Страница 12: ... the spring force make sure that the spring arm with the lamp device can come to rest in any desired position A hole 1 is located at the position marked by detail X Position the spring arm 2 with the lamp device approximately 10 above horizontal Insert Allan key width 5 included in the scope of supply into the hole 1 If the spring arm drops the spring force is too low Rotate the adjustment screw t...

Страница 13: ... Allan key width 5 in the hole 6 If the spring arm drops the spring force is too low Rotate the adjustment screw to the left counter clockwise in the direction If the spring arm rises the spring force is too high Rotate the adjustment screw to the right clockwise in the direction If the spring arm with the lamp device cannot come to rest in any desired position after the spring force has been adju...

Страница 14: ...des a little tighter using an open ended or ring wrench SW14 or loosen these nuts while at the same time holding the screw slot 3 in each case with a screw driver Once the adjustment has been made place seal 4 correctly on the housing lower part Replace the upper part of the lamp housing so that the seal lies cleanly around the circumference Using a round object drawing pin or small screw driver l...

Страница 15: ...d s ecurity clip 2 and turn shade holder 3 down Carefully remove the halogen bulb 4 22 8 24V 50W from its socket change and replace Replace shadeholder and the disk bearer onto the bottom part of the lamp housing and rotate the quick release fasteners 1 clockwise Do not touch the halogen bulb with naked hands Remove stains with a clean cloth and alcohol since otherwise these can burn into the glas...

Страница 16: ...ase you notice a reduced light quality because of a dull splinter protection disk it may be necessary to change the disk To change the splinter protection disk proceed as follows Turn off the lamp Rotate the three quick release fasteners 1 half a turn anticlockwise so that the corresponding disk bearer can be removed from the bottom part of the lamp housing Remove filter disk 4 as described at poi...

Страница 17: ...alternating current rated voltage rated current frequency Hertz fuse class of protection 5 2 Environmental conditions Operation Min Max Temperature 10 C 40 C Relative atmospheric humidity 30 75 Air pressure 700 hPa 1060 hPa Transport Storage Min Max Temperature 10 C 50 C Relative atmospheric humidity 20 90 Air pressure 700 hPa 1060 hPa ...

Страница 18: ... 42 EEC for medical products of the European Community s Council 7 Disposal The OT lamp Quintaflex R96 does not contain any danger goods The components of the OT lamp should be properly disposed at the end of their shelf life Make sure that the materials are separated accordingly For disposal proceed as follows PC boards should be submitted to an appropriate recycling The rest of the components sh...

Страница 19: ...Quintaflex R96 Ceiling Wall Model Dr Mach Lamps and Engineering 59250001 Edition 01 21 02 2002 Bak Page 19 18 8 Spare parts Power supply with 24V ceiling model ...

Страница 20: ...complete 25060002 20 21 5 Glass reflector 10060201 22 5 Flange 23060208 23 5 Cooling body 23060214 24 1 Handle bow complete 25040001 25 1 Cardan bow complete 25110001 26 27 1 Inner ring for needle bearing in part 25 26 23110202 28 29 30 5 Socket 67320005 31 5 Halogen bulb 22 8V 50W 67100201 32 5 Tracer pin 25080202 33 34 1 Focussing gear complete 25050001 35 1 Housing 25050201 36 1 Cap 25054001 37...

Страница 21: ...Quintaflex R96 Ceiling Wall Model Dr Mach Lamps and Engineering 59250001 Edition 01 21 02 2002 Bak Page 21 18 51 52 15 Nut M3 DIN 934 65332002 53 1 Acrylic glass box for transformer 300VA 67610101 ...