TWISTER
®
Softwareversion 1.19
Page 44 of 46
Version 1.19-4 / 2010-05-31
11 Serial number design
TW
-
xxx
-
12
-
2007
Twister
Consecutive No.
Calendar week
Year of manufacture
12 Disposal of waste
The TWISTER
®
stimulator has to be disposed according to the locally valid instructions and
standards for electronic scrap.
13 Cleaning
The TWISTER
®
device has to be cleaned with a moist cloth or disinfection by wipe.
Cleaning agents must not be used!
14 Service and repair
The intended use is based on the compliance with the required service intervals.
Service intervals:
Prior to each operation:
Visual examination for damages of the housing and accessories and functional test is
required.
Annually:
Measurement of the leakage currents and the resistance of the protective earth wire
according to IEC601-1 (safety check) has to be done by the manufacturer.
Содержание TWISTER