Version 1.19-4 / 2010-05-31
4 Functional principals and operation
The maximum voltage is +/- 200 Volts.
The operational indicator (READY) at the front panel of the TWISTER
®
shows, that the TWISTER
®
is ready to deliver stimulation pulses.
The
electrode output
is twice isolated against power supply and the control unit, consisting
of the trigger output, footswitch input and USB link and thus potential free.
Only those
electrodes
approved by the Dr. Langer Medical GmbH are allowed to be
connected to the electrode output.
The stimulation current density must not exceed 2 mA / cm²
.
Energy limitation:
TWISTER
®
monitors the energy delivered to the electrodes. The energy is measured at a
1000 Ohm resistance and limited to a maximum of 50 mJ. The exceeding will be signalized
by the software.