background image

 

Ultra Evac Technical Manual  

26 

2/8/2013 

GD-99013 Rev. D 

Company Information

 

DMS has been helping healthcare facilities responsibly manage infectious fluid waste since 1997.  Starting 
out  as  the  premier  manufacturer  for  fluid  waste  management  systems  in  the  country,  we  now  have 
installations in leading healthcare facilities nation-wide. 

DMS  was  founded  in  1995.    Products  are  manufactured  and  shipped  from  a  centralized  facility  and 
supported  by  a  nation-wide  sales  force.      The  company‟s  principals  include  individuals  with  extensive 
backgrounds  in  product  development,  management,  production,  and  sales  of  medical/surgical  products.  
Jim Dunn, a successful inventor who served as an operating room nurse for 20 years, helped design DMS 
products.   

 

Contact Information 

 

Dornoch Medical Systems, Inc. 

 

S h i p p i n g   A d d r e s s :  

200 NW Parkway 

Riverside, MO 64150 

 

O f f i c e   C o r r e s p o n d e n c e :  

P.O. Box 681656 

Riverside, MO 64168 

 

Toll free: 1-888-466-6633 

Phone: 1-816-505-2226 

Fax: 1-816-505-1050 

Web

www.dornoch.com

 

 

Содержание UL-EV100

Страница 1: ...Ultra Evac Technical Manual 2 8 2013 GD 99013 Rev D Transposal Ultra Evacuation Unit Technical Manual Model No UL EV100...

Страница 2: ...Ultra Evac Technical Manual 2 8 2013 GD 99013 Rev D...

Страница 3: ...Ultra Evac Technical Manual 2 8 2013 GD 99013 Rev D...

Страница 4: ...Enzyme Bottles 10 Reprocessing Cycle Descriptions 10 Performing Verify Bleach Cycle 10 Unit Information and Error Messages 11 Periodic Maintenance 11 Troubleshooting Guide 12 DMS Technical Service In...

Страница 5: ...ra Evac Unit Dimensions and Features Dim inch cm 5 Figure 2 TP DL2800 Single Use Lid 6 Figure 3 UL CL500 Single Use Manifold 6 Figure 4 Ultra Fluid Cart Connected to the Ultra Evac Unit 6 Figure 5 Cou...

Страница 6: ...or Use Manual This manual is designed to instruct users in the correct operation of the evac as well as safety considerations associated with the unit Technical Service Manual This manual includes ope...

Страница 7: ...operating maintenance and safety information Intended Use The intended use of the evac is to process the carts after they have collected surgical fluids The evac will empty clean and rinse the reusab...

Страница 8: ...ce personnel should open any of the access covers on this equipment User operation does not require access to these areas ALWAYS use the handle to move the cart DO NOT push or pull the cart by graspin...

Страница 9: ...nding Installation Fixed Size inch cm 20 51 W x 16 41 D x 28 71 H Weight lbs kg 100 46 Utilities Power 120VAC 60Hz 15amp Water 3 4 Hot and Cold 5 0gpm 18 9lpm 55psi Hot Water Temperature 100 115oF 38...

Страница 10: ...hnical Manual 5 2 8 2013 GD 99013 Rev D Figure 1 Ultra Evac Unit Dimensions and Features Dim inch cm Power Indicator Evac Coupler Waste Hose Power Inlet Coupler Hoses Hot Water Inlet Leveling Legs 33...

Страница 11: ...o uses two single use lids or manifolds and the Ultra Quad uses four lids or manifolds New lids are pressed on to the top of cart reservoirs or new manifolds are inserted into the housings On carts wi...

Страница 12: ...Controls The Ultra Fluid Carts high tech control system guides users through unit operation while monitoring key functions Bleach Cycle Monitoring Continuous automatic monitoring of the bleach cycle...

Страница 13: ...d using a Verification Log Kit TP VL100 Continuous automatic monitoring of the bleach cycle is accomplished with electronic sensors within the evac If an error is detected the message shall be reporte...

Страница 14: ...out of the receiver Place the receiver in the storage unit CAUTION ALWAYS press down gently on the handle prior to removing the coupler Failure to do so could cause damage to the equipment Reprocessi...

Страница 15: ...rvoir that are not coming off during a wash Let the fluid soak for at least one hour for best results Rewash Cycle will attempt to clean the selected reservoir s again using wash water and enzyme The...

Страница 16: ...s message indicates that the Enzyme solution has run out The unit alarms at the end of the cycle when the reservoirs have completed the cleaning process Replace the enzyme with a new gallon Valve Erro...

Страница 17: ...em Reservoirs are Not Washing Possible Solution 1 Verify the water to the unit is turned on 2 Check for any crimps in the black water line to the coupler on the back of the evac 3 Call DMS at 1 888 46...

Страница 18: ...and patent information Operating Instructions Located on the back of unit near the power cord The label is used to instruct the user to the manuals as well as inform them of the power outlet requireme...

Страница 19: ...of consumables associated with the evac WARNING Use only Dornoch approved components and enzyme DO NOT modify any component or accessory UL EZ200 Enzymatic Cleaner TP VL100 Verification Log and Test S...

Страница 20: ...risk damage to or loss of the installation items that may be sent with the unit Contact customer service if unpacking the unit is necessary Site Preparation A DMS Installation Technician is responsibl...

Страница 21: ...pressure 55 psi or higher at the unit Hot supply pipe should run up or down the wall to the left side of the space where the machine will be installed ending 48 122cm to 58 148cm from the floor Attac...

Страница 22: ...Ultra Evac Technical Manual 17 2 8 2013 GD 99013 Rev D Figure 9 Ultra Evac Unit Architectural Drawing...

Страница 23: ...rial Always unplug the unit from the wall electrical receptacle before moving equipment Wipe down the unit with an approved contact disinfectant after each use Equipment Storage Temperature range for...

Страница 24: ...y result in damage to or malfunction of the system or even personal injury Service Warnings WARNINGS Modification of this equipment is not allowed When performing any service or maintenance procedures...

Страница 25: ...anel has a ground wire attached to it Inspections Regularly visually inspect the unit for any signs of damaged or leaking components If any problems are found contact Customer Service at 1 888 466 663...

Страница 26: ...iagrams Ultra Cart Wash Ultra Cart Waste Hot Water Supply Master Valve Wash Valve Clog Buster Valve Evac Wash Manifold Evacuation Pump Chemical Pump Flow Sensor Bleach Reservoir Enzyme Reservoir 2 Way...

Страница 27: ...Ultra Evac Technical Manual 22 2 8 2013 GD 99013 Rev D Figure 12 Ultra Evac Unit Wiring Diagram...

Страница 28: ...Ultra Evac Technical Manual 23 2 8 2013 GD 99013 Rev D...

Страница 29: ...Ultra Evac Technical Manual 24 2 8 2013 GD 99013 Rev D Figure 13 Ultra Evac Unit Board Schematic...

Страница 30: ...maintenance parts kits can be ordered directly from DMS by calling 1 888 466 6633 Table 3 Preventative Maintenance Schedule Year End Required Parts Procedure 1 2 3 4 Part Quantity Annual PM Kit X X X...

Страница 31: ...d shipped from a centralized facility and supported by a nation wide sales force The company s principals include individuals with extensive backgrounds in product development management production an...

Страница 32: ...HE CONTRARY DMS DOES NOT WARRANT ANY OF THE FOLLOWING THAT MAY BE SHIPPED WITH OR AN INTEGRAL PART OF ANY TRANSPOSAL PRODUCT OR SUPPLIES REQUIRED TO OPERATE OR INSTALL ANY TRANSPOSAL PRODUCT I ITEMS A...

Страница 33: ...OUT OF OR RELATED TO THIS LIMITED WARRANTY WILL BE LITIGATED IN COURTS HAVING SITUS IN KANSAS CITY MISSOURI PURCHASER HEREBY CONSENTS AND SUBMITS TO THE JURISDICTION OF ANY COURT LOCATED WITHIN MISSOU...

Отзывы: