Analog Output
Figure 3.10:
Channel(s) configuration window, Analog Output CW
The
Analog Output
channel type creates analog pulse sequences. Each numbered channel corresponds to the same
number analog channel on the console. Pulse sequences have different parameters depending on the channel
Mode
,
which can be
Continuous
,
Square
or
Sine
. The
Channels Configuration
window contains the following elements for an
analog channel.
• The
CW (Continuous wave)
channel mode (Fig.
) allows the creation of a continuous analog signal. The fol-
lowing elements appear in the
Sequence Options
box.
1. The
Starting Delay
(Fig.
) defines the time between the activation of the pulse sequence and the begin-
ning of the signal.
2. The
Time ON
(Fig.
) defines the length of time the continuous signal is active. Should the time chosen be
0, the signal will continue until the pulse sequence is stopped manually.
3. The
Voltage
(Fig.
) defines the voltage of the continuous signal, in volts. The signal cannot go beyond
±
4.75 V.
4. The
Total Duration
(Fig.
) shows the total expected duration of the pulse sequence. Should the duration
be infinite, the box will display
∞
. If there is an error in parameter selection, this box will display
N/A
.
Chapter 3. Doric Neuroscience Studio