development), the measurement sequence when a high TTL signal (
>
4 V) is detected, and will stop when
a low TTL signal (
<
0.4 V) is detected.
This source can either be
Manual
, or from one of the four
Digital I/O
channels. The
Trigger Mode
can be
either
Triggered
or
Gated
.
c) The
Pre-processing
(Fig.
) filters define the high and low-pass frequency cutoff for electrical signal re-
ceived by the headstage.
3. The
LED options
(Fig.
) are all parameters used to control the light source of the cannula connected to the
Fi-Wi
headstage
.
a) The
Channel options
(Fig.
) are used to control the LED mode and current.
i. The
Mode
(Fig.
) allows the selection of the pulse sequence mode. At time of writing, only
Square
mode is available.
ii. The
Maximum current
(Fig.
) defines the current sent to the cannula LED. For proper function of the
cannula, the current should always be greater than 10 mA.
iii. The
Baseline
(Fig.
) leaves a small offset to the current sent to the LED. It is reccomended to use a
small offset, as a complete shut-down of the LED will induce a spike in the electrical acquisition signal.
b) The
Sequence options
(Fig.
) box is where LED pulse sequence parameters are defined.
i. The
Starting Delay
(Fig.
) sets the delay (in hh:mm:ss:zzz format) before the first pulse.
ii. The
Frequency/Period
(Fig.
) sets the frequency (in Hz) or period (in ms) for the pulse sequence.
For example, a signal at 10 Hz (frequency) will output one pulse every 100 ms (period), whereas a pulse
sequence at 0.5 Hz (frequency) will output one pulse every 2000 ms (period).
iii. The
Time ON/Duty Cycle
(Fig.
) sets the time (in ms) or the duty cycle (in %) for each pulse. The
Time
ON
must be lower than (1/fre0.005 ms, while the
Duty cycle
must be below 100 %. These
squares will appear red should an impossible
Frequency
/
time ON
be selected.
iv. The
Smoothing
option is used to change the pulse slope in square pulse sequences. The
Edit Edges
button opens the
Smoothing Edge(s)
window (Fig.
Figure 3.3:
Light Source Smoothing Edge(s) Window
A. The
Rise Time
box is used to define the duration to rise from 0 to the pulse maximum.
B. The
Plateau Time
box is used to defined the duration the pulse is at its maximum value.
C. The
Fall Time
box is used to define the duration to descend from the pulse maximum to 0.
D. The
Pulse Graph
displays the pulse shape.
E. The
Active Time
box displays the total duration of the pulse. While the
Smoothing
option is active,
the
Time ON
is fixed at this value.
(need to include section
v. The
Pulses per sequence
(Fig.
) sets the number of pulses per sequence. If it is set to 0, the pulse will
be repeated indefinitely.
vi. The
Number of sequences
(Fig.
) sets the number of times that the sequence will be repeated. If it is
set to 0, the sequence will be repeated indefinitely.
vii. The
Delay between sequences
(Fig.
) sets the delay (in hh:mm:ss:zzz format) between each sequence
if the
Number of Sequences
is greater than 1.
viii. The
Total Duration
(Fig.
) displays the total time of the experiment. The different values can be
Inf
for infinite, a valid time value or
Err
if the
Time ON
value is greater than 1/frequency.
c) The
Preview
box shows a preview of the pulse sequence.
Chapter 3. Doric Neuroscience Studio
14