33
Donaldson Company, Inc.
Troubleshooting
Problem
Probable Cause
Remedy
Blower fan and motor do
not start
Improper motor wire size
Rewire using the correct wire gauge as specified by
national and local codes.
Not wired correctly
Check and correct motor wiring for supply voltage.
See motor manufacturer's wiring diagram. Follow
wiring diagram and the National Electric Code.
Unit not wired for available
voltage
Correct wiring for proper supply voltage.
Input circuit down
Check power supply to motor circuit on all leads.
Electrical supply circuit down
Check power supply circuit for proper voltage.
Check for fuse or circuit breaker fault. Replace as
necessary.
Insufficient airflow
Fan rotation backwards
Proper fan rotation is clockwise from the top of the
unit. The fan can be viewed through the back of the
motor. See Preliminary Start-Up Check.
Fan exhaust area restricted
Check fan exhaust area for obstructions. Remove
material or debris. Adjust damper flow control.
Filters need replacement
Remove and replace using genuine Donaldson
replacement filter packs. See Filter Replacement.
Lack of compressed air
See Rating and Specification Information for
compressed air supply requirements.
Pulse cleaning not energized
Use a voltmeter to check the solenoid valves in the
control panel. Check pneumatic lines for kinks or
obstructions.
Pulse control timer board out of
adjustment
See Solid-State Timer Installation.
Compressed air supply wet or
oily
Bleed compressed air supply. Check for debris,
valve wear, or diaphragm failure by removing the
diaphragm cover on the pulse valves. Check for
solenoid leaks or damage. If pulse valves or solenoid
valves and tubing are damaged, replace.
Pulse valves leaking
compressed air
Lock out all electrical power to the unit and bleed
the compressed air supply. Check for debris, valve
wear, pneumatic tubing fault, or diaphragm failure by
removing the diaphragm cover on the pulse valves.
Check for solenoid leaks or damage. If pulse valves
or solenoid valves and tubing are damaged, replace.