Using the refrigerator
21
7.2
Starting up the device
!
I
•
At the first starting up of the device or after switching it off for a long time the
display of the digital thermometer (fig.
2
, page 4) shows the current ambient
temperature alternating with the error message “E6”.
•
The alarm LED will light up but no audible alarm will sound.
•
As soon as the inner temperature reaches the operating temperature range
(2 °C – 8 °C) the current temperature is shown.
➤
Reset the thermometer to avoid a wrong maximum temperature value.
➤
Press the
button for 5 seconds.
7.3
Display backlight in battery mode
The thermometer display has a blue backlight to make the temperature values easier
to read. The display backlight works continuously while the device is connected to
the mains.
In battery mode the display backlight will switch off after 2 minutes.
➤
To activate the display backlight for 3 seconds, briefly press the
button.
7.4
Temperature values
The thermometer stores the values of the minimum and maximum inner temperature
since the last reset.
To reset these values:
➤
Press the
button for 5 seconds.
CAUTION!
Do not store medicines and medicinal products before the E6 error
code has disappeared and the operating temperature range has been
reached.
NOTE
The temperature values are shown in °C. This cannot be changed.
HC302D-502D-IO-WEST.book Seite 21 Freitag, 17. März 2017 2:15 14
Содержание HC302D
Страница 2: ......
Страница 3: ...HC302D HC502D 3 2 1 5 7 8 6 HC302DFS HC502DFS HC302DBIU HC502DBIU 3 3 4 1...
Страница 4: ...HC302D HC502D 4 5 6 7 8 4 3 1 2 2 1 2 1 1 1 2 2 2 A D C B 3...
Страница 5: ...HC302D HC502D 5 50 mm 4 50mm 5...
Страница 6: ...HC302D HC502D 6 min 100 6 1 2 3 7...
Страница 7: ...HC302D HC502D 7 4 1 1 3 2 8 1 2 3 4 1 2 3 4 9...
Страница 8: ...HC302D HC502D 8 C D A B 0 820 422 FS 450 BIU 440 540 580 HC302D a...
Страница 9: ...HC302D HC502D 9 HC502D 948 486 FS 494 BIU 484 552 592 b...