C L 1 U S E R S G U I D E
D O D T E C H N O L O G I E S , I N C
Rev 10.07
Page 34
8
Optic Comm Failure
Electrical problem
Remove optic block cover and make
sure all wires going to the optics are
fully inserted. Contact DOD
Technnologies support (815) 788-
5200 if these steps don't correct the
problem.
9
All Pumps Disabled
Flow Blocked and/or
Unstable Inlet/Exhaust
When blockage is fixed, go to Setup
Menu and enable pumps. If problem
persists, contact DOD (815) 788-
5200.
10
Flow Fluctuation Fault
Flow Blocked and/or
Unstable Inlet/Exhaust
When blockage is fixed, go to Setup
Menu and enable pumps. If problem
persists, contact DOD (815) 788-
5200.
11
Idle Timeout
Out of Analysis Beyond
Timer
Re-enter Analysis
12
Low Background Fault
Wrong tape or optic
problem
Re-adjust optics - make sure fresh
tape is inserted and gate is closed.
Then go to service menu-optics and
press start. If these steps don't
work, contact DOD DOD
Technnologies support (815)788-
5200
13
Purge Option Fault
Tubing Blocked
Find block in sample line
14
Optic Auto Cal Failure
Optics Dirty
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. Also, re-adjust optics -
make sure fresh tape is inserted and
gate is closed. Then go to service
menu-optics and press start.
33
SD Card Fault
No SD Card or Card Full
If a SD card is installed remove and
re-insert
34
Analyzing
Analysis mode started
N/A
35
Chemlogic Tape Low
ChemLogic tape nearing
end
Replace tape
36
Pump 1 Disabled
Flow Blocked and/or
Unstable Inlet/Exhaust
Go to Setup Menu and enable pump.
If problem persists, contact DOD
(815) 788-5200.
37
Pump 2 Disabled
Flow Blocked and/or
Unstable Inlet/Exhaust
Go to Setup Menu and enable pump.
If problem persists, contact DOD
(815) 788-5200.
38
Large Flow Fluctuation
Flow Blocked and/or
Unstable Inlet/Exhaust
When blockage is fixed, go to Setup
Menu and enable pumps. If problem
persists, contact DOD (815) 788-
5200.
39
High Flow Fault
Optics Orifice Dirty
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. If problem persists, contact
DOD (815) 788-5200.
40
Alarm Level 1
Gas Detected
After gas leak is found and
corrected, reset alarms
41
Alarm Level 2
Gas Detected
After gas leak is found and
corrected, reset alarms