6.4 End of line particulate filters
End of line (point of detection) particulate filters which protect the CL1 from damage are
required for most gases. Table 6.1 details the type of filter required for each gas – Filters must
be replaced on a regular basis as shown in the table. Filter orientation is not critical in either
application.
Gas
Description
Suggested Replacement
DOD Filter Part #
AsH3
Arsine
6 Months
780248
B2H6
Diborane
6 Months
780248
GeH4
Germane
6 Months
780248
H2SE
Hydrogen Selenide
6 Months
780248
PH3
Phosphine
6 Months
780248
SiH4
Silane
6 Months
780248
TBA
Tertiary-Butyl-Arsine
6 Months
780248
H2S
Hydrogen Sulfide
6 Months
780248
HCL
Hydrogen Chloride
1 Month (membrane)
60009 (Housing) 60010 (membrane)
HF
Hydrogen Flouride
1 Month (membrane)
60009 (Housing) 60010 (membrane)
BF3
Boron Triflouride
1 Month (membrane)
60009 (Housing) 60010 (membrane)
HBR
Hydrogen Bromide
1 Month (membrane)
60009 (Housing) 60010 (membrane)
COCL2
Phosgene
6 Months
780248
Diisocyanates
NO FILTER
N/A
Table 6.1
6.5 Flow adjustment
The pumps in the CL1 automatically adjust to keep a constant flow to the system. No manual
adjustment is necessary.
6.6 Micro Secure Digital (SD) Card replacement
See Appendix F for detailed information on the contents of the SD Card.
To install a Micro SD card: Align its 8-pin gold edge connector down, facing the front of the
CL1 unit as shown in Figure 6.3; then carefully push it all the way into the Memory slot.
Ensure that it clicks into place. To remove the Micro SD card: Push down on the top of the
card gently to release the spring. The card pops up for removal. Make sure to format the SD
card before first use whenever a new card is purchased.