BEFORE USING THE DEVICE, PLEASE READ THE FOLLOWING
INSTRUCTIONS COMPLETELY AND CAREFULLY. CORRECT
APPLICATION IS VITAL TO THE PROPER FUNCTIONING OF THE DEVICE.
INTENDED USER PROFILE:
The intended user should be a licensed medical professional, the patient or
the patient’s caregiver. The user should be able to read, understand and be
physically capable of performing the directions, warnings and cautions in
the information for use.
INTENDED USE / INDICATIONS:
The Aircast AirSelect Standard walker is designed to provide support and
immobilization to the lower leg, ankle and foot. It may be suitable for use
following stable fractures of the foot and/or ankle and lateral ankle sprains.
PERFORMANCE CHARACTERISTICS:
Soft-good/ semi-rigid designed to restrict motion through elastic or
semi-rigid construction.
Providing immobilization or controlled movement of the limb or body
segment.
CONTRAINDICATIONS:
This device is contraindicated for unstable fractures of the lower leg, ankle
and foot. Do not use if you are allergic to any of the materials contained
within this product.
WARNINGS AND PRECAUTIONS:
•
This device is to be used under the supervision of a healthcare
professional. The determination of when to apply the brace and the
frequency and duration of use should be strictly at the discretion of the
treating physician. Always consult with your physician before making
changes to the brace.
•
Do not use this device on patient’s incapable of communicating
physical discomfort.
•
Do not over-inflate aircells. Improper aircell inflation may cause
significant skin irritation, especially in people with diabetic neuropathy.
Reduce aircell pressure with any visual skin changes or reported
discomfort.
•
Patient’s without sensation (i.e. post-op anesthesia, neuropathies, etc.)
should be monitored frequently for skin damage, skin irritation or
wound management.
•
Use caution when walking on slippery or wet surfaces to avoid injury.
•
Do not use over open wounds.
•
Do not over tighten straps. This may result in reduction of blood flow or
sensation.
•
Do not modify the device or use the device other than intended.
•
At high altitudes the aircells will expand beyond their optimal level.
Adjust aircell support by tightening or loosening straps.
•
Do not use this device if it is damaged and/or the packaging has been
opened.
•
If pain, swelling, changes in sensation or other unusual reactions
occur while using this product, consult your medical professional
immediately.
NOTE:
Contact manufacturer and competent authority in case of a serious
incident arising due to usage of this device.
Use a sneaker or street shoe with a ShoeLift™ or equivalent on uninjured
foot while ambulating to compensate for leg length changes while wearing
a fracture walking brace.
F
注記:機能や強度、耐久性、快適性を最適化できるよう最先端技術によりあらゆる努力が払われていますが、本製
品の使用により負傷が予防されるという保証はまったくありません。
2
31
ENGLISH
E
3) エアセルサポートの調整
STANDARD
歩行ブレースのエアセルは空気を抜いた状態で支給され
ます。
• 損傷側から膨張をはじめます。ブレースの右側にあるセレクタを適切
な番号の方向に回します(図D)。1番は右側のエアセルを膨張させ、
2番は左側のエアセルを膨張させます。エアセルの膨張が完了したら、
「ロック」の位置に変更してください。誤って空気が抜けるのを防ぎ
ます。
• エアセルを膨張させるには、ブレースの右側上部にある
大型
膨張ポンプの
上の小さな穴を指で塞ぎます(図E)。エアセルがぴったりするまで膨張させ
ます。
• もう一方のエアセルを膨張させる場合、セレクタを適切な番号の方向に回
し、大型膨張ポンプを押し、解放します。エアセルがぴったりするまで膨張さ
せます。両方のエアセルの膨張が完了したら、SELECTORを「ロック」の位置
に回転して固定させてください。
• 膨張しすぎた場合にはエアセルを収縮させます。
4) エアセルの収縮
• エアセルを収縮する場合、ブレースの右側にあるセレクタを適切なエアセ
ル番号の方向に回します(図D)。
• ブレースの右側上部にある
小型
リリースボタン(図F)を押し、解放します。
• 圧力が適度になるまで収縮させます。
注意:
再装着の際にぴったりと装着できるように、
ブレースを取り外す前に
エ
アセルを収縮させます。
注意:
圧が高いほど歩行時のサポートは強く、座ったり横になったりする場合
は低めの圧力がより快適です。圧力はまたストラップを締める/緩める
ことでも調整できます。
注意:
より幅広のふくらはぎ用に調整する場合は、ギプスカッターやギプスは
さみでシェルを切ってください。
洗浄時の注意:
フォームライナーは低刺激性洗剤を溶かした水(30°C)で手洗いし、空気乾燥さ
せてください。乾燥機に入れたり他の熱源を使って乾燥させたりしないでくだ
さい。
保証:
DJO, LLCは、材料や工程上の瑕疵に対しては、販売日から6か月間に限り、本体
および付属品の全部または一部を修理もしくは交換します。
天然ゴムラテックス不使用
単独患者使用限定
材料成分
シェル:ポリプロピレン
正面パネル:高密度ポリエチレン
ストラップ:ナイロン
膨張ポンプ:超可塑性ポリウレタン樹脂
ライナー:ナイロン/ポリウレタン/ポリエステル(外側/発泡スチロール/内部メ
ッシュ)
結合剤:ポリエステル
外側靴底/中敷:エチレン酢酸ビニール/ゴム
セレクタダイヤル:ナイロン