12
ENGLISH
System Alarms
Alarm Reset
To reset any alarm, press the button and take steps to correct the alarm if
necessary . Pressing the button does not fix the alarm but silences it until the problem
is fixed . If you wish to turn the device completely off, you must press the power button
once to silence alarm and a second time to power down the device .
S Mode Operation
When powered on, the VenaFlow Elite System will default to standard rapid inflation
mode . If a slower inflation is desired, select the S mode operation
button . Once
selected, both cuffs will inflate simultaneously once per minute . Both cuffs will inflate
and compress for a total of 10 seconds . While in S mode the icon
will appear in
the top center of the graphical display . The device will remain in S mode until user
presses the S mode operation button again . This button allows user to toggle between
standard rapid inflation mode and S mode .
Locking the System in Standard Rapid or S Mode
You have the ability to lock the system in either standard rapid mode or S mode . Once
a mode is locked, it will stay in that mode regardless of powering the device on or off .
Once locked, it is not possible to toggle modes until the system is unlocked .
To lock the system in standard rapid inflation mode:
First
make sure the device is in standard rapid inflation mode . The S
mode icon should NOT appear on the display . Next, press and
hold the S mode button
for no less than 5 seconds . After 5
seconds, a lock will appear at the top center of the display .
This lock icon indicates that the system is locked in standard
rapid mode .
To lock in S mode:
First select S mode and make sure the
S mode icon
appears at the top center of the display .
Press and hold the S mode button
again for no less than
5 seconds . After 5 seconds, a lock will appear next to the S
mode icon
at the top center of the display . This lock icon
next to the S mode icon indicates that the system is locked in S
mode .
To unlock either mode:
Press and hold the S mode button
for no less than 5
seconds . After 5 seconds, the lock icon will disappear .
Содержание 30BI-S
Страница 31: ......