Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The VenaFlow Elite is intended for use in the electromagnetic environment specified below .
The customer or the user of the VenaFlow Elite should assure that it is used in such an
environment .
Immunity Test
IEC 60601 Test
Level
Compliance Level Electromagnetic Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood,
concrete or ceramic tile . If floors
are covered with synthetic
material, the relative humidity
should be at least 30% .
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for power
supply lines
±1kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment .
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common
mode
±1kV differential
mode
±2kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment .
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
) for 0,5 cycle
40%
U
T
(60% dip in
U
T
) for 5 cycles
70%
U
T
(30% dip in
U
T
) for 25 cycles
<5%
U
T
(>95% dip in
U
T
) for 5 sec
<5%
U
T
(>95% dip
in
U
T
) for 0,5 cycle
40%
U
T
(60% dip in
U
T
) for 5 cycles
70%
U
T
(30% dip in
U
T
) for 25 cycles
<5%
U
T
(>95% dip
in
U
T
) for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment . If the user of
the VenaFlow Elite requires
continued operation during
power mains interruptions,
it is recommended that the
VenaFlow Elite be powered
from an uninterrupted power
supply or a battery .
Power
Frequency
(50/60Hz)
Magnetic Fields
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment .
NOTE:
U
T
is the a .c mains voltage prior to application of the test level .