11
ENGLISH
Pump Operation
1) To turn the device on, press . The graphical display, green pump indicator lights
and green light above button will turn on .
2) Once device is powered on, system will immediately enter cuff detection mode
which means the system will be detecting whether or not there are cuffs attached .
Display will read “Detecting Cuffs” and will display percentage of detection
complete . Once it reaches 100%, the display will then transition to the standard
display screen .
3) Any changes to cuff configuration requires that the device be powered off and
restarted in order to detect the new configuration .
4) To turn the device off, press . The graphical display, green pump indicator lights
and green light above button will turn off .
5) To shut down the power completely, disconnect the power plug by removing it
from the AC socket .
Single Leg Operation
1) Cuffs should be attached prior to powering on the device . Once powered on, the
device will automatically detect if one or two cuffs are attached .
2) Any changes to cuff configuration requires that the device be powered off and
restarted in order to detect the new configuration .
3) Either port may be used for single leg operation . The system will automatically
detect which one is in use .
4) Single and dual leg operation are available in both standard rapid mode and S
mode operation .
Patient Compliance Counter Reset
To reset the patient compliance counter, press for 1 second and release . The hours,
minutes and seconds will reset .