background image

7

1-800-634-7328    DirectSupply.com

Installation Of Head and Foot Panels 

Refer to Bed Panel Installation Sheet for more

Tools required: #2Phillips screwdriver

Head Panel Installation:

1.  Head panel has label on back.

2.  Lay panel on soft surface front face down.

3.  Head panel brackets are longer of two sizes (Figure 5).

4.  Install head panel brackets with four screws provided. 

Bracket offset should face up (Figure 6). Do not tighten 

until after fitting mounting brackets into bed sockets.

5.  Go to final assembly below.

Frame-Mounted Foot Panel Installation:

1.  Foot panel does not have label on back. Foot panel 

will have staff control electronics, if staff control was 

ordered.

2.  Lay panel on soft surface front face down.

3.  Foot panel brackets are shorter of two sizes (Figure 5).

4.  Install the foot panel brackets with four screws 

provided. Bracket offset should face up (Figure 6) Do 

not tighten until after fitting mounting brackets into bed 

sockets.

5.  Go to final assembly below.

Final Assembly Step for Head Panel and Frame-
Mounted Foot Panel:

1.  Insert head and foot panel as shown with mounting 

hardware facing head of the bed  

(Figure 7).

2.  Tighten all screws with panels in place.

3.  Panels should slip in and out of their sockets easily and 

securely.

IV Sockets

Direct Supply Panacea 1000-FLEX Bed Platform has four 

IV sockets as standard equipment. There are two sockets 

at head end of bed and two at foot end (Figure 8). The IV 

sockets are used with most ½" IV poles.

Figure 5

Head Panel 

Bracket-10.5"

Foot Panel 

Bracket-6.9"

Offset facing up

Figure 6

Figure 7

Figure 8

IV Socket

Содержание Panacea 1000-FLEX BED

Страница 1: ...uipment Furnishings Please read this entire guide carefully and keep it for future reference This guide will provide you with instructions warnings warranty informa tion and other important informatio...

Страница 2: ...wall mounted outlet in good working order with no cracks or chips The plug should fit tightly into the outlet The ground pin should be intact and secure Do not use power strips extension cords or mult...

Страница 3: ...tient by a qualified clinician or medical provider to ensure maximum safety of the resident patient The assessment should be conducted within the context of and in compliance with the state and federa...

Страница 4: ...as should be used no closer than 30 cm 12 inches to any part of the Direct Supply Panacea 1000 FLEX bed including cables specified by the manufacturer Otherwise degradation of the performance of this...

Страница 5: ...C to 70 C Relative Humidity 10 to 100 Atmospheric Pressure 500 hPa to 1060 hPa Mechanical Specifications Description Dimensions Overall Dimensions 86 9 L x 35 0 W x 8 75 H Weight Total without Panels...

Страница 6: ...igure 3 2 Remove the screws using the 2 Phillips screwdriver do not discard the screws 3 Position and insert the mattress stops in the holes provided Figure 4 Reinstall the screws removed in step 2 4...

Страница 7: ...soft surface front face down 3 Foot panel brackets are shorter of two sizes Figure 5 4 Install the foot panel brackets with four screws provided Bracket offset should face up Figure 6 Do not tighten...

Страница 8: ...ant Route one cable under the control box and plug into pendant port Route the second cable under the control box and connect to the hand pendant Figure 9 4 Route the two cables under the main frame t...

Страница 9: ...from contacting floor This will help prevent cord damage when transporting bed CAUTION If a power cord is damaged remove bed from service until repairs are performed WARNING Possible Injury Before lea...

Страница 10: ...foot frame by compatible ID labels on left side Figure 16 2 Lay frame halves on soft surface feet side down Place packing material under head frame to make assembly easier Figure 17 Ensure the split f...

Страница 11: ...amage to the actuators 3 Install the back actuator and knee actuator on the head frame and foot frame as shown using the actuator pin and clip Figure 19 4 Insert head frame into foot frame and line up...

Страница 12: ...23 Figure 24 1 Line up back and knee actuators and insert actuator pin then add clip to lock in place Figure 22 and 23 Ensure the washers are used on the back actuator Figure 24 Figure 22 Split Frame...

Страница 13: ...hen route behind cable bundle Figure 25 and 26 3 Insert back actuator cable in port 1 and insert hi lo actuator head end cable in port 3 Figure 27 4 Secure cables with cable clamp Figure 28 5 Route po...

Страница 14: ...tuators and control box contain no serviceable parts Overload protection is achieved via current sensing technology which will automatically reset If any bed function fails to operate check to see if...

Страница 15: ...wer cord to frame Tape remaining cables in their original position It is very important to keep actuator cables routed along frame in original factory installed positions If this step is not followed...

Страница 16: ...he bed is plugged into power source and control box green light is illuminated 4 Thermal overload condition This should self correct in 2 minutes or less Do not surpass 10 duty cycle 2 min on 18 min o...

Страница 17: ...yclable items properly To prevent possible harm to the environment or human health from uncontrolled waste disposal recycle it responsibly to promote the sustainable reuse of material resources CAUTIO...

Страница 18: ...to disconnect from bed resulting in loss of control of bed 4 Take extra care when moving bed remember do not move bed while occupied by resident patient on any sloped surface Moving a bed on a sloped...

Страница 19: ...1 800 634 7328 for instructions 3 Install the trapeze assembly onto the trapeze adapter according to the trapeze manufacturer s instructions 4 Operate the bed from its highest through lowest position...

Страница 20: ...FLEX bed is intended for use in an electromagnetic environment for professional healthcare in which radiated RF disturbances are controlled The customer or the user of the Direct Supply Panacea 1000 F...

Страница 21: ...61000 4 5 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth Mains power quality should be that of a typical professional heal...

Страница 22: ...kHz Portable and mobile RF communications equipment should be used no closer to any part of the Direct Supply Panacea 1000 FLEX bed including cables than the recommended separation distance calculate...

Страница 23: ...e to someone else or at the end of the applicable limited warranty period identified below whichever is earlier Warranty Period Parts Only Welds 15 years Mechanical Components 3 years Electronics 3 ye...

Страница 24: ...Distributed by Direct Supply Manufacturing Inc 7301 W Champions Way Milwaukee WI 53223 2021 Direct Supply Manufacturing Inc All Rights Reserved Direct Supply Panacea and all associated logos are the...