4

 

WARNING: Possible Injury Or Death.

 Do not use any 

Panacea

®

 assist device until you verify it is designed specifically 

for the Direct Supply Panacea 1000-FLEX bed and it is locked in 
place. Failure to lock assist devices may result in injury or death.

 

WARNING: Possible Injury or Death. 

NEVER permit more 

than (1) person in /on the bed at any time. NEVER exceed the 
weight capacity of this bed. The weight capacity of this bed is 450 
lbs (204 kg) including accessories and options (not all accessories 
and options may be used at the same time). Subtract residents/
patients weight from beds capacity 450 lbs (204 kg) to determine 
allowable weight for mattress, bed panels, accessories, and 
optional equipment. See Accessories Section for accessory 
weights.

 

WARNING: Possible Injury Or Death

. Do not use any 

replacement parts not manufactured, marketed, or provided 
by Direct Supply on any Direct Supply Panacea bed. Use of 
unapproved replacement parts may result in injury or death.

 

WARNING: Possible Injury Or Death.

 If using accessories 

not manufactured, marketed, or provided by Direct Supply for 
Direct Supply Panacea beds, consult with the manufacturer for 
compatibility and limitations prior to use. Failure to do so may 
result in injury or death.

 

CAUTION:

 Radio transmitting equipment, cellular phones, or 

similar electronic devices, used in proximity of bed, may affect 
bed performance.

 

CAUTION:

 Particular precaution must be considered during 

exposure to strong emission sources such as High Frequency 
surgical equipment and   similar devices. Do not route High 
Frequency cables on or near device. If in doubt, contact a qualified 
technician.

NOTE:

 Electromagnetic disturbances may contribute to the 

improper operation of medical devices.  In the case of a bed, 
it may affect the performance of the bed movement when a 
function is selected.  Do not place the bed in a location that is 
known to have high electromagnetic fields.

 

WARNING:

 Use of this equipment adjacent to or stacked with 

other equipment should be avoided because it could result in 
improper operation. If such use is necessary, this equipment and 
the other equipment should be observed to verify that they are 
operating normally.

 

WARNING:

 Use of accessories, transducers, and cables other 

than those specified or provided by the manufacturer of this 
equipment could result in increased electromagnetic emissions or 
decreased electromagnetic immunity of this equipment and result 
in improper operation.

 

WARNING: 

Portable RF communications equipment (including 

peripherals such as antenna cables and external antennas) should 
be used no closer than 30 cm (12 inches) to any part of the Direct 
Supply Panacea 1000-FLEX bed, including cables specified by the 
manufacturer. Otherwise, degradation of the performance of this 
equipment could result.

NOTE:

 The EMISSIONS characteristics of this equipment make it 

suitable for use in industrial areas and hospitals (CISPR 11 class 
A). If it is used in a residential environment (for which CISPR 
11 class B is normally required) this equipment might not offer 
adequate protection to radio-frequency communication services. 
The user might need to take mitigation measures, such as 
relocating or re-orienting the equipment.

 

CAUTION:

 This device complies with EMC requirements of 

IEC 60601-1-2. Radio transmitting equipment, cellular phones or 
similar electronic devices, used in proximity of bed, may affect 
bed performance.

 

CAUTION:

 In the absence of line voltage, all bed adjustments 

can be achieved via an externally available backup battery (Direct 
Supply Panacea P/N N211).

 

WARNING: Possible Shock Hazard.

 If the bed power cord has 

a hospital-grade, 3-prong grounding plug, grounding reliability can 
only be achieved when equipment is connected to an equivalent 
receptacle marked hospital-grade.

 

WARNING: Possible Injury.

 Sitting on or sliding over the 

Mattress Side Stop can cause skin injury if the bed mattress 
compresses to less than 1.5 inches under user’s weight.

 

WARNING: Possible Injury. 

Risk of entanglement or injury 

may occur if mattress used with Mattress Side Stop does not fill 
the entire width between stops or which compresses to less than 
1.5 inches under user’s weight.

 

CAUTION: 

The Direct Supply Panacea 1000-FLEX bed is 

intended for use within institutional healthcare environments 2, 
3, and 5 (ie: skilled nursing, transitional care, rehabilitational care, 
assisted living). Direct Supply recommends compliance to the 
regulations and guidelines specific to your locality.

Panacea assist devices are designed to be used with specific 
Panacea bed frames. Once mounted on the appropriate bed frame, 
assist devices become an integral part of the Panacea bed system. 
As such, these mounted assist devices, meet the IEC 60601-2-52 
Bed Standard, within normal use conditions, of the appropriate bed 
to which they are mounted. 

Assemble, install, utilize, and maintain Panacea Assist Devices as 
described in their accompanying Panacea User/ Service Manual, 
which also lists the bed type or model with which these devices 
may be used. 

This product is only one part of your healthcare bed system. 
Proper combinations of mattress, head/foot panels and assist 
devices are needed to minimize the risk of entrapment. For more 
information, call Customer Care, 1-800-634-7328.

There are no known contraindications for this device.

Important Precautions

 (cont.)

 

Содержание Panacea 1000-FLEX BED

Страница 1: ...uipment Furnishings Please read this entire guide carefully and keep it for future reference This guide will provide you with instructions warnings warranty informa tion and other important informatio...

Страница 2: ...wall mounted outlet in good working order with no cracks or chips The plug should fit tightly into the outlet The ground pin should be intact and secure Do not use power strips extension cords or mult...

Страница 3: ...tient by a qualified clinician or medical provider to ensure maximum safety of the resident patient The assessment should be conducted within the context of and in compliance with the state and federa...

Страница 4: ...as should be used no closer than 30 cm 12 inches to any part of the Direct Supply Panacea 1000 FLEX bed including cables specified by the manufacturer Otherwise degradation of the performance of this...

Страница 5: ...C to 70 C Relative Humidity 10 to 100 Atmospheric Pressure 500 hPa to 1060 hPa Mechanical Specifications Description Dimensions Overall Dimensions 86 9 L x 35 0 W x 8 75 H Weight Total without Panels...

Страница 6: ...igure 3 2 Remove the screws using the 2 Phillips screwdriver do not discard the screws 3 Position and insert the mattress stops in the holes provided Figure 4 Reinstall the screws removed in step 2 4...

Страница 7: ...soft surface front face down 3 Foot panel brackets are shorter of two sizes Figure 5 4 Install the foot panel brackets with four screws provided Bracket offset should face up Figure 6 Do not tighten...

Страница 8: ...ant Route one cable under the control box and plug into pendant port Route the second cable under the control box and connect to the hand pendant Figure 9 4 Route the two cables under the main frame t...

Страница 9: ...from contacting floor This will help prevent cord damage when transporting bed CAUTION If a power cord is damaged remove bed from service until repairs are performed WARNING Possible Injury Before lea...

Страница 10: ...foot frame by compatible ID labels on left side Figure 16 2 Lay frame halves on soft surface feet side down Place packing material under head frame to make assembly easier Figure 17 Ensure the split f...

Страница 11: ...amage to the actuators 3 Install the back actuator and knee actuator on the head frame and foot frame as shown using the actuator pin and clip Figure 19 4 Insert head frame into foot frame and line up...

Страница 12: ...23 Figure 24 1 Line up back and knee actuators and insert actuator pin then add clip to lock in place Figure 22 and 23 Ensure the washers are used on the back actuator Figure 24 Figure 22 Split Frame...

Страница 13: ...hen route behind cable bundle Figure 25 and 26 3 Insert back actuator cable in port 1 and insert hi lo actuator head end cable in port 3 Figure 27 4 Secure cables with cable clamp Figure 28 5 Route po...

Страница 14: ...tuators and control box contain no serviceable parts Overload protection is achieved via current sensing technology which will automatically reset If any bed function fails to operate check to see if...

Страница 15: ...wer cord to frame Tape remaining cables in their original position It is very important to keep actuator cables routed along frame in original factory installed positions If this step is not followed...

Страница 16: ...he bed is plugged into power source and control box green light is illuminated 4 Thermal overload condition This should self correct in 2 minutes or less Do not surpass 10 duty cycle 2 min on 18 min o...

Страница 17: ...yclable items properly To prevent possible harm to the environment or human health from uncontrolled waste disposal recycle it responsibly to promote the sustainable reuse of material resources CAUTIO...

Страница 18: ...to disconnect from bed resulting in loss of control of bed 4 Take extra care when moving bed remember do not move bed while occupied by resident patient on any sloped surface Moving a bed on a sloped...

Страница 19: ...1 800 634 7328 for instructions 3 Install the trapeze assembly onto the trapeze adapter according to the trapeze manufacturer s instructions 4 Operate the bed from its highest through lowest position...

Страница 20: ...FLEX bed is intended for use in an electromagnetic environment for professional healthcare in which radiated RF disturbances are controlled The customer or the user of the Direct Supply Panacea 1000 F...

Страница 21: ...61000 4 5 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth 0 5 kV 1 kV line s to line s 0 5 kV 1 kV 2 kV line s to earth Mains power quality should be that of a typical professional heal...

Страница 22: ...kHz Portable and mobile RF communications equipment should be used no closer to any part of the Direct Supply Panacea 1000 FLEX bed including cables than the recommended separation distance calculate...

Страница 23: ...e to someone else or at the end of the applicable limited warranty period identified below whichever is earlier Warranty Period Parts Only Welds 15 years Mechanical Components 3 years Electronics 3 ye...

Страница 24: ...Distributed by Direct Supply Manufacturing Inc 7301 W Champions Way Milwaukee WI 53223 2021 Direct Supply Manufacturing Inc All Rights Reserved Direct Supply Panacea and all associated logos are the...

Отзывы: