Guidance and Manufacturer’s Declaration - Electromagnetic Immunity:
The Direct Supply Panacea 1000-FLEX bed is intended for use in the electromagnetic environment (for professional
healthcare) specified below. The customer or the user of the Direct Supply Panacea 1000-FLEX bed should assure that it
is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment-Guidance (for
professional healthcare environment)
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact: ±8 kV
Air: ±2 kV, ±4 kV,
±8 kV, ±15 kV
Contact: ±8 kV
Air: ±2 kV, ±4 kV,
±8 kV, ±15 kV
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
Not applicable
Mains power quality should be that of a typical
professional healthcare environment.
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV line(s)
to line(s)
±0.5 kV, ±1 kV, ±2 kV
line(s) to earth
±0.5 kV, ±1 kV line(s)
to line(s)
±0.5 kV, ±1 kV, ±2 kV
line(s) to earth
Mains power quality should be that of a typical
professional healthcare environment.
Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Voltage Dips:
0%
U
T
; 0,5 cycle
0%
U
T
; 1 cycle
70%
U
T
; 25/30 cycles
Voltage Interruptions:
0%
U
T
; 250/300 cycle
Voltage Dips:
0%
U
T
; 0,5 cycle
0%
U
T
; 1 cycle
70%
U
T
; 25 cycles
Voltage Interruptions:
0%
U
T
; 250 cycle
Mains power quality should be that of a typical
professional healthcare environment. If the user
of the Direct Supply Panacea 1000-FLEX bed
requires continued operation during power mains
interruptions, it is recommended that the Direct
Supply Panacea 1000-FLEX bed be powered from
an uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz / 60 Hz
The Direct Supply Panacea 1000-FLEX bed power
frequency magnetic fields should be at levels
characteristic of a typical location in a typical
professional healthcare environment.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.