background image

 

 

 

August 2003 

Rev. 2 

Page 8

 

 

© 

2003 Dionex 

 

  

Please note:

 

Column identification is supported for two columns simultaneously. 

  

Please note:

 

To store and read out the column-specific information, use the 
corresponding commands and properties in Chromeleon. For more 
information, refer to the 

Chromeleon online Help

.  

Column identification is not supported in stand-alone control. 

5. Configuring the Column Switching Valve 

  

Please note:

 

In the example below, two eluent preconditioners are used. For 
information about how to install the eluent preconditioner, refer to the 
Installation Instructions shipped with the units. 

 

Remove the plastic cap that protects the 
column switching valve during shipment. 

For a typical example for the connections on 
the valve, see the photo above. 

Position 1: From autosampler/pump

Position 6:  
To column// 
preconditioner inlet B

Position 2:  
To column// 
preconditioner inlet A

Position 3:  
From column outlet A

Position 4: To detector 

 

Position 5:  
From column outlet B

Содержание TCC-100

Страница 1: ...August 2003 Rev 2 Page 1 2003 Dionex TCC 100 Thermostatted Column Compartment Quick Installation Guide ...

Страница 2: ...o not connect the unit to the mains during this period If it is obvious that condensation did not evaporate completely continue to warm up the unit which is not connected to the mains until the condensation is completely gone Place the instrument on a firm and level surface that is free of vibration Make sure that the surface is resistant to solvents Avoid locations with extreme changes in tempera...

Страница 3: ...ooled to preset temperature set point To increase or decrease the value press the Plus or Minus key until the display starts flashing You can change the set point in increments of 0 1 C as long as the display is flashing To change the temperature value more quickly hold down the key The display stops flashing when you release the key and then displays the actual temperature again To interrupt oper...

Страница 4: ...n compartment driver via the Setup program on the software CD provided in the accessories kit Insert the CD into the CD drive of your Chromeleon server PC Double click Setup exe in the root directory Click Next to have the Setup program check which Chromeleon version is installed on your PC Follow the on screen instructions as they appear to install the driver If a version is found for which no dr...

Страница 5: ... the other supported Windows operating systems we recommend using a USB connection whenever possible For information about the TCP IP setup please refer to the Operating Instructions for the TCC 100 Afterward install the TCC 100 in the Chromeleon Server Configuration program Follow the instructions in the corresponding sections of the Operating Instructions for the instrument or refer to the Chrom...

Страница 6: ... tilts downward and the fan stops automatically if temperature control is enabled Tighten the column clips into place using a screw driver Press the column into the column clips To close the front panel tilt it upwards until you hear the locking mechanism engage The fan restarts automatically Please note For precise temperature control make sure that the front panel is closed properly Push Press ...

Страница 7: ...chip card and read it out whenever you want Clip the memory chip card onto the column as shown above To do so you can open the expanding rivet by removing the rivet head from the body Take the chip card Make sure that the Dionex logo is facing you Insert the chip card with the Dionex logo facing up into one of the column identification slots You may now close the front panel door Expanding rivet E...

Страница 8: ...Column Switching Valve Please note In the example below two eluent preconditioners are used For information about how to install the eluent preconditioner refer to the Installation Instructions shipped with the units Remove the plastic cap that protects the column switching valve during shipment For a typical example for the connections on the valve see the photo above Position 1 From autosampler ...

Страница 9: ... position 6 to the inlet of either a column or eluent preconditioner here to the eluent preconditioner installed in the bottom position in the column chamber Connect the capillary from valve position 2 to the inlet of either the second column or second preconditioner here to the eluent preconditioner installed in the top position Connect the capillary to the detector inlet to valve position 4 ...

Страница 10: ...olumn into the column clips In this example the column is installed directly on the eluent preconditioner Install a capillary between valve position 3 and the outlet of the corresponding column here to the column installed in the top position Please note If valve position 2 was used for the column inlet valve position 3 must be used for the column outlet Connect a capillary to valve position 5 and...

Страница 11: ...umn inlet valve position 5 must be used for the column outlet Connect the preconditioner outlet o the inlet of the relevant column here to the inlet of the column installed in the bottom position Configuration of the column switching valve and the installation of both columns is now complete Please note In the above picture the Column Identification System is used for both columns ...

Отзывы: