![Dini Argeo DGT1P Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/dini-argeo/dgt1p/dgt1p_user-manual_2505733024.webp)
24
Optimized layout for A4 print.
888888
i
f.mode
A
888888
Setup
B
ConfiG
pre.Cal
1
1
deCi
man.Cal
Calib
1
8
4
f.mode
A
888888
Setup
B
ConfiG
pre.Cal
1
1
0.Calib
5
TECH_MAN_ENG_DGT1P
MENU ACCESS:
SAVING THE PARAMETERS:
Press the
key during the start-
up procedure.
Press the
key several times, until the display shows
SAVE?
. Press the
key to confirm.
Complete menu
on pages
24 - 25
Manual calibration
If you know the number of ADC converter points for a known weight (for example if you want to copy the calibration from one transmitter
to another) the calibration points can be entered manually:
1.
The display shows
mod.pnt
, and then the last calibration point.
2.
Using the and keys, select the calibration point you want to enter / change (from
0
to
8
).
Press the
key to confirm.
3.
The display shows
weiGht
, use the , and keys to enter the weight value.
Press the
key to confirm.
4.
The display shows
points
, use the , and keys to enter the converter points value.
Press the
key to confirm.
Quick zero calibration (pre-tare reset)
Tp 0
Tp n
ADC
Kg
Tp 0
Tp n
ADC
Kg
Содержание DGT1P
Страница 1: ...www diniargeo com DGT1P Digital weight transmitter ENGLISH USER MANUAL...
Страница 2: ......
Страница 6: ...6 Optimized layout for A4 print TECH_MAN_ENG_DGT1P...
Страница 51: ......