21
Optimized layout for A4 print.
1
3
2
pre.Cal
Setup
1
B
ConfiG
1
deCi
div
ranGe1
theo.Ca
u.m.
Calib
1
3
4
7
2
4
Cel.sen
Cel.CAp
dead.ld
4
kno.wGt
888888
f.mode
A
TECH_MAN_ENG_DGT1P
LEGEND:
Indicates repeated pressing of
the
key.
Parameter visible only
under certain conditions.
Parameter or menu subject
to approval.
Default value of the parameter.
Theoretical calibration
CALIBRATION PROCEDURE:
1.
Set the calibration parameters:
deCi
= Number of decimals.
u.m
.
= Unit of measurement (
kG
,
G
,
t
,
lb
).
div
= Minimum division.
ranGe1
= Maximum range.
2.
Set the cell data:
sen.Cel
= Cell sensitivity (given by the average mV/V value of cells).
Cel.Cap
= Total capacity of the cells (given by the sum of the capacities of each cell).
3.
Enter the weight value of the structure in the
dead.ld
parameter. If you do not know this value, enter “0”.
4.
If the structure contains a quantity of material whose weight value is known (e.g. full silo), enter this value in the
KNo.WGT
parameter.
5.
Application of theoretical calibration:
Press the key to exit the calibration menu. The display shows
th.Cal?
. Press the
key to confirm the use of the theoretical
calibration, or the key to cancel.
Содержание DGT1P