22
is as follows:
Key
Lower
display
Upper
display
meaning
【
Fxx
】
[
F01
]
[
d , , , * * *
]
In the weighing display state, press and hold
the [FXX] key, and the lower row will
flash.If the lower row displays [
F00
].Then
enter the password above the administrator
level, log in first, and then perform this
step again
Press and
hold
[
F25
]
[
,A-in,,,
]
Press and hold [↑] continuously to quickly
locate F25.
[input]
[
F25
]
[
04--00.00
]
On the left, the current external input
accurate current number is displayed, and on
the right, the current value measured by the
instrument input port is displayed. Increase
the output current value by pressing
【↑ 】
and decrease the current value by pressing
【↓ 】
[input]
[
F25
]
[
16--00.00
]
Press the [←] →] key to adjust the input
current. Take
16mA as an example. Press [↑]
to increase the output current value and [↓]
to decrease the current value. Through the
adjustment of two points, the accuracy of
analog input port is calibrated.
[input]
[
F25
]
[
,A-in,,,
] Adjustment completed.
Table 7.4 Specific steps of analog input port calibration and setting
7.6 Settings of panel display content
The panel display is double-row digital tube display, the specific content
of which can be set freely, and the F parameters involved are F20 and F21. DSP1
in parameter F20 corresponds to the upper row main display, KP1 corresponds to
the upper row auxiliary display, DSP2 in parameter F21 corresponds to the lower
row main display, and KP2 corresponds to the upper row auxiliary display. All four
parameters can be set to the following:
serial
number
parameter
Show content
serial
number
parameter
Show content
1
000
The main display
shows 0/ the
auxiliary display
does not show.
5
100
Process executor
status
2
001
Strain P01
6
101
Cumulative times