RFLS-35-dat-4.6

The sensor can be mounted in a tank or at medium inlets. After setting to the level of the given media the sensor does not react 
to the current of flowing medium.

Installation of the scanner for reliable checking of the level of a fluid with foam on the surface. Sensitivity of the sensor can be 
set to detect the liquid interface with foam. After a drop in the fluid level, the sensor does not react to coatings of foam on the 
electrode. 

the LED indicator “STATE”  

is not lit

the LED indicator “STATE” is lit

In the case of 

side wall mounting

. it is necessary to avoid long fitting tubes, where sensed medium could adhere (image on the 

left). We recommend mounting the sensor so that the whole measuring electrode is inside the tank (image on the right).

It  is  recommended  to  install  sensors  in  a 
horizontal pipe 

inclined from the side.

or  adhering  remnants  of  viscous 
media at the bottom of the pipe.

!

!

Upon vertical installation of the sensor in 
a pipe, pay attention to potential formation 
of air pockets, 