The sensor can be mounted in a tank or at medium inlets. After setting to the level of the given media the sensor does not react
to the current of flowing medium.
Installation of the scanner for reliable checking of the level of a fluid with foam on the surface. Sensitivity of the sensor can be
set to detect the liquid interface with foam. After a drop in the fluid level, the sensor does not react to coatings of foam on the
electrode.
the LED indicator “STATE”
is not lit
the LED indicator “STATE” is lit
In the case of
side wall mounting
. it is necessary to avoid long fitting tubes, where sensed medium could adhere (image on the
left). We recommend mounting the sensor so that the whole measuring electrode is inside the tank (image on the right).
It is recommended to install sensors in a
horizontal pipe
inclined from the side.
or adhering remnants of viscous
media at the bottom of the pipe.
!
!
Upon vertical installation of the sensor in
a pipe, pay attention to potential formation
of air pockets,