OPERATOR’S MANUAL
GAC® 2500 AGRI
11001-1626B-201405
TROUBLESHOOTING
/
81
ERROR
CODE
ERROR
PROBABLE CAUSE
CORRECTIVE ACTION
1
Empty Cell Measurement
Out of Spec
Moisture or dirt buildup in cell.
Dump grain and verify cell is clean and free of
grain in and surrounding the cell.
2
Empty Cell Weight Out of
Spec
Empty cell measurement is out of tolerance
Dump grain and verify cell is clean and free of
grain in and surrounding the cell.
3
No Products Installed
No product calibrations were found on the machine.
Install product calibration files.
4
Fill Motor Jammed
The motor has stalled while loading grain from a
possible obstruction.
Clear blockage from cell.
5
Invalid Grain Calibration File
An invalid calibration file was selected.
Re-install the selected calibration file.
6
Moisture Too High
Measured product is above the moisture upper limit
of the calibration.
Verify sample filled the cell.
7
Moisture Too Low
Measured product is below the moisture lower limit
of the calibration.
Verify sample filled the cell.
8
Instrument Low Temp Limit
Exceeded
Instrument temperature is less than the allowed
limit of 0 degrees C.
Move unit to a warmer environment or allow to
warm up.
9
Grain High Temp Limit
Exceeded
Grain temperature has exceeded the unit
specification to analyze grain.
Allow grain to cool and then re-analyze.
10
Grain Low Temp Limit
Exceeded
Grain temperature has exceeded the unit
specification to analyze grain.
Allow grain to warm and then re-analyze.
11
Sample Weight Too High
Test weight is greater than the range specified for
the calibration selected.
Use a more representative sample and
re-analyze grain.
Verify correct grain calibration is selected.
12
Sample Weight Too Low
Test weight is lower than the range specified for the
calibration selected.
Use a more representative sample and
re-analyze grain.
Verify correct grain calibration is selected.
13
No Communication
An internal communication failure has occurred and
communication has been lost.
Power unit off and turn back on to reset.
14
Instrument High Temp Limit
Exceeded
Instrument temperature is higher than the allowed
limit of 45 degrees C.
Check unit ventilation or move to a cooler
environment.
15
Unit to Grain Differential
The difference between grain temperature and
machine temperature has exceeded the unit
specification to analyze grain.
Allow grain and machine temperature to equalize
and then re-analyze.
16
Internal Power Supply Out of
Spec
Unit internal voltage is out of specification.
Service is required. Contact DICKEY-john Tech
Support at 1-800-637-3302.
17
Unable to Predict Moisture
A corrupt instrument calibration file or other
unexpected error.
Dump sample and re-analyze.
19
Instrument Needs Updated
Firmware in the instrument is out of date for this
application.
Update device with new firmware.
21
File I/O Error
An unexpected file i/o error has occurred.
An obstruction has occurred in the cell. Blockage
must be cleared before testing can resume.
22
RF interference detected.
Unable to compute moisture
Outside 2-way radio interference or cell board
communication interference
Remove outside radio interference. If problem
persists, contact DICKEY-john Technical Support
at 1-800-637-3302.
50
Weight Measurement Device
Error
An error has occurred with the load cell.
Power unit off and turn back on to reset.
Service unit if failure continues.
51
Invalid Password
A service function with an invalid password was
attempted.
If service is required, contact DICKEY-john
Technical Support at 1-800-637-3302 for
assistance on how to obtain the password and
proper service procedure.
52
Date/Time Incorrect
The programmed machine date is less than
expected.
Program the instrument date and time. If problem
persists, then the battery backup for the internal
clock may need to be replaced. Contact
DICKEY-john Technical Support at
1-800-637-3302 for service.
53
Cold Sample Moisture Too
High
The sample is too high in moisture to accurately
read at its current temperature.
Warm the sample above the low temperature
threshold specified in the product calibration file
and rerun the measurement.
55
Dump Motor Timeout
The dump mechanism does not appear to be
closing.
Check that the dump motor optical sensor is
clean and there is no interference in the dump
mechanism. Press the Green button to retry. If
error persists, contact technical support at
1-800-637-3302.
Содержание GAC 2500-AGRI
Страница 4: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 IV...
Страница 20: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 16 NAVIGATION...
Страница 68: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 64 GRAIN CALIBRATIONS...
Страница 70: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 66 DIAGNOSTICS...
Страница 82: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 78 MAINTENANCE...
Страница 84: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 80 TROUBLESHOOTING...