CLEANING THE TEMPERATURE SENSOR PROBE
IMPORTANT: Any particles in front of the sensor’s optics can affect
measurement performance. Therefore it is crucial to
sufficiently clean the sensor. For excessive dust and/or
foreign material buildup on or around the temperature
sensor, it is recommended the instrument be returned to
DICKEY-john Service or authorized service center.
The temperature sensor may require cleaning due to dust buildup and/or
foreign material that has collected around the sensor that could potentially
cause temperature error readings during analysis. It is important the
temperature sensor is visually inspected and cleaned each time a regular
maintenance check of the instrument is conducted.
Tools required for cleaning the temperature sensor probe:
1. One of the below cotton swabs:
– Standard 6” long cotton swab (i.e. McMaster-Carr p/n 7074T12) for
flush sensor
– 6” long, tapered tip cotton swab (i.e. McMaster-Carr p/n 71035T54)
for embedded sensor
Figure 72
Recommended Cotton Swab for Flush Sensor
Figure 73
Recommended Cotton Swab for Embedded Sensor
2. 99% Isopropyl Alcohol
Figure 74
99% Isopropyl Alcohol
3. 10” flat head screw driver (only for instruments with flush sensor/brush
bracket.
Figure 75
10” Flat Head Screw Driver
Содержание GAC 2500-AGRI
Страница 4: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 IV...
Страница 20: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 16 NAVIGATION...
Страница 68: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 64 GRAIN CALIBRATIONS...
Страница 70: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 66 DIAGNOSTICS...
Страница 82: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 78 MAINTENANCE...
Страница 84: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 80 TROUBLESHOOTING...