Abacus junior 5
User’s Manual
57
7.5.2.4.1 Multi-user mode
If the analyzer is in multi-user mode, a login
screen will be displayed during start-up,
asking for a
user ID
and a
password
.
After the User ID is entered and confirmed
with the OK key, the user name
corresponding to the ID appears in the
bottom line. If the user name was correct,
enter the
password
, and confirm with the
function key. If the password was
correct, the analyzer continues initialization
and is ready to work.
In multi-user mode the Shut down menu is
changed;
Logout menu point appears.
SHUT DOWN (1)
A logged in user who has finished working
with the instrument should select SHUT
DOWN (1).
When logout is used instead of shutdown,
the log-in screen will be displayed for the
next user, and pneumatics will not be shut
down.