2.3.2 Turning the Instrument OFF
The instrument should never be switched off by simply pressing the power button on
the rear panel. Doing so may result in erroneous operation during later use. The
instrument uses different kinds of solutions, one of which is the so-called diluent. This
liquid is an isotonic saline solution containing salt. Would it not be washed out of
special units of the instrument, or would chambers not be filled with this solution may
lead to dust condensation, or salt build up.
Therefore, always follow the instructions below when switching the instrument off.
Select EXIT key
. The following screen is displayed.
EXIT
SHUT DOWN (1)
The software will prompt you for confirmation. The analyzer will perform a priming
cycle, filling the chamber to avoid dust and salt build-up. The screen will prompt the
user again to power off accompanied by a continuous beep.