7.3 Quality Control
By analyzing control materials, day-to-day reproducibility can be monitored. Both
target values and acceptable ranges for each parameter can be specified for different
QC levels herein.
The
Abacus junior 5
provides six different Quality Control levels. You can
configure up six individual reference sheets for each control material (e.g. low,
normal and high control blood). QC measurement results will be saved to the
selected level, as indicated at the top right corner.
NOTE:
Target values of the control material should be set only once, at
the beginning of the QC measurements. Resetting parameters
deletes previous QC results of the actual level.
CAUTION! Any change in the QC material setting deletes previous QC results.
It is strongly recommended to print results prior to changes.
From UTILITIES you can access Quality
Control Menu
UTILITIES
QUALITY CONTROL (3)
This menu allows entering and monitoring of
data related to the QC procedure.
Use commercially available control material.
The preparation and analysis process is the
same as with patient samples.
UTILITIES
QUALITY CONTROL (3)
SET QC LEVEL (5)
Select the level you wish to use.
The active level is displayed in the top right
corner on QC related screens.
UTILITIES
QUALITY CONTROL (3)
SET QC REFERENCE (1)
Both target values and acceptable ranges can
be specified. Only parameters displayed on
these screens are utilized. Modify displayed
values using the numerical keyboard. OK key
accepts data. Use the page down function key
to view additional parameters. Setting 0.0
disables QC of parameter.