Error! Use the Home tab to apply Címsor 1 to the text that you want to appear
here.
51
If Waste is high, it should be disposed of properly (see next section for instructions).
MAINTENANCE
REAGENT STATUS
Volume
Set the volume of the containers used
(Values in ml). If volume of a reagent is set to
0 (zero), software will not keep track of
consumption.
.
Accept
saves your changes.
Cancel
discards changes and returns to reagent status screen
7.5.1. How to empty waste container
Software counts volume of waste, and gives warning message when the waste tank is
close to its maximum capacity.
Empty the waste tank when this warning message appears. See next Section for
neutralization steps.
7.5.2. Neutralization of Waste
Waste contains human origin substances representing biohazard. These substances are
representing potential danger to environment. For this reason, safe handling of the waste
liquid is very important
Neutralization of biohazard waste:
Put 2 ml per liter of
hypochlorite
solution into the waste. Close the cap and shake the
container.
After 1 hour you can dispose of the Waste liquid into the drain.