INSTALLATION
14
2.
INSTALLATION
2.1.
General information
This chapter provides instructions for the installation of
A
A
b
b
a
a
c
c
u
u
s
s
J
J
u
u
n
n
i
i
o
o
r
r
3
3
0
0
hematology analyzer. The procedures described below must be followed correctly to
ensure proper operation and service. Please carefully read and follow all instructions
in this
User’s Manual
before operating the analyzer.
This hematology analyzer is a precision instrument: handle with care. Dropping or
other improper handling of the instrument will disturb calibrated mechanic and
electronic components and/or cause damage.
CAUTION: Always handle the instrument with care.
2.2.
Environmental factors
Operate
A
A
b
b
a
a
c
c
u
u
s
s
J
J
u
u
n
n
i
i
o
o
r
r
3
3
0
0
within the ambient temperature range of 15-
30°C and
relative humidity of 45% .. 85
%. The optimum operating temperature is 25°C.
Avoid using the instrument in areas of extreme high or low temperatures or where it
is exposed to direct sunlight. If kept at a temperature less than 10°C, the instrument
should be allowed to sit for an hour at the correct room temperature before use.
Reagents should be stored at a temperature range of 18-
30°C.
Place the instrument in a well-ventilated location. Do not place it near potentially
interfering devices capable of emitting radio frequencies (e.g. radio or television
receiver, radars, centrifuge, X-ray devices, fans, etc.).
Operation at an altitude over 3000 meters (9000 ft) is not recommended, because the
throughput will be degraded.
Instrument is safe for transient voltages to INSTALLATION CATEGORY II and
POLLUTION DEGREE 2.
Environmental and electrical characteristics provide accuracy and precision of the
instrument and maintain a high level of operational safety for lab personnel.
2.2.1. Electrical requirements
A
A
b
b
a
a
c
c
u
u
s
s
J
J
u
u
n
n
i
i
o
o
r
r
3
3
0
0
comes with an approved power cord, appropriate for your power
system. Proper use of the appropriate power cord assures adequate grounding of the
system.
WARNING: Improper grounding of the analyzer bypasses important safety
features and may result in electrical hazard.