1. Before using the handpiece, check that charging is complete and the handpiece is ready for use.
2. Check that the disposable gun tips and heat chamber are clean and undamaged before each use.
3. As plunger is completely inserted, do not force to pull trigger which may result in device
damage.
4. Do not re-use needles. After obturation, remove disposable gun tips with wrench completely.
5. After each use, remove all visible soil from Gutta Percha Obturator in the heat chamber with
scrubbing brush. Melt the residual at 160
and clean it all.
6. Use one pellet of Gutta Percha Obturator only. Many pellets can damage the device.
7. To reduce the injury of patient or operator, always place the safety cap on during use.
8. Gun tip should be tightened moderately so that there is no leakage or flow out of Gutta Percha
Obturator. Over-tightening the tip may lead to damage to the tip and/or handpiece. The tip can
be adjusted and bent to the desired degree by the supplied wrench. Do not bend the needle by
force.
9. To prevent injury, when the gun tip is activated, avoid contact with the patient and operator.
When the gun tip needs to be removed after obturation, wait at least three (3) minutes until the
tip has been cooled down.
10. After use, push the plunger backwards to the start-position. There is a possibility that the
plunger might not move back because the remaining filling material is frozen inside. In this
case, operate the device and move plunger back slowly.