F. Obturating (Backfill)
- Fit the master Gutta Percha cone that is dipped in sealer and take a radiograph to ensure that it is
filling well.
- Increase the preset temperature of Dia-Gun, so Gutta Percha Obturator is heated.
- Using Dia-Gun s tip, make contact with the upper part of Gutta Percha Point that is already filled
in the root canal, transfer heat for 3 to 5 seconds.
- Operate the trigger slowly, flow the melted Gutta Percha Obturator to the root canal and continue
flowing while withdrawing the needle from the canal.
- Use an Endodontic hand plugger to compact obturation material.
- Repeat as needed until canal is filled without voids and take a radiograph.
G. Powering Off
- Press the power button for one (1) second to turn the device off.
- Power turns off automatically in four and a half (4.5) minutes when the device is not in use.