8
Handpiece
A. Power On/Off and Temperature Setting Button :
Press the button for one (1) second to turn the device On/Off.
Press the button shortly to increase or decrease preset temperatures (160
, 180
, 200
).
Power turns off automatically in four and a half (4.5) minutes when the device is not in use. This
automatic time-out feature prevents overheating.
B. Display :
Indicates setting and heating temperature.
C. Loading Slot :
Place Gutta Percha Obturator to the heat chamber.
D. Plunger :
Push the Gutta Percha Obturator into the heat chamber until it snaps securely into place.
Plunger O-rings could be worn by the repetition use and caused the back-flow, please change O-
rings by the Multipurpose Wrench.
E. Trigger :
Lightly squeeze the trigger to achieve a flow of warm Gutta Percha Obturator.
F. Heat Chamber with safety cap :
Swivel 360 to adjust needle s direction. Soft safety cap can
reduce the possibility of injury during use. Always steam autoclave the safety cap before use.
G. Disposable gun tips (23G) :
Connect the tip tightly to heat chamber and bend or remove using
the multipurpose wrench.
H. Multipurpose Wrench :
After the needle is placed properly, bend and tighten it to the desired
degree using the wrench. After use, remove the needle safely.
I. Rechargeable Battery Pack :
Inside handpiece.
J. Reset Switch :
Before using the handpiece of Dia-Gun for the first time, you must press the
Reset Switch that is located at the bottom of the handpiece once. When the handpiece is off, its
display is blank. Pressing the Reset Switch once will light up the display. Should this device not
work properly because of internal program error, please power off Reset Switch and power it on
after one (1) minute.