Diablo 5000B Real-Time Gas Analyzer Getting Started Manual
9
Setting up the Hardware
Connecting sample lines
Caution:
Particulates and condensed liquids or droplets in the sample stream can
plug the 30-
μm sample orifice, which will then need to be replaced. It is the
responsibility of the user to ensure that their sampling system is designed
to eliminate particulate and liquid prior to the sample gas stream reaching
the RTGA sampling tee.
The sample input and output connections are standard 1/16-inch Swagelok fittings. By using 1/16-
inch tubing, the dwell time of the sample line can be minimized without introducing a large pressure
drop. The sample input is connected to the top of the interface and the vent or process return is
connected to the bottom of the interface. To make a new fitting connection, make sure the tube is
fully inserted into the fitting body. Finger-tighten the fitting and then tighten ¾ of a turn with a 5/16-
inch wrench on the compression nut and a 5/16-inch wrench on the adapter body.
Figure 2: A photograph of the RTGA interface with the sampling tee identified.
If necessary, user-supplied fittings can be used to adapt process lines to the 1/16 inch required for
the interface. Heating the sample lines is the responsibility of the user.