background image

 

5.  TROUBLESHOOTING 

 

Page 27 

© 2007 DH Instruments, a Fluke Company 

5

5

.

.

 

 

T

T

R

R

O

O

U

U

B

B

L

L

E

E

S

S

H

H

O

O

O

O

T

T

I

I

N

N

G

G

 

 

GPC1 is a sophisticated pressuring generating and adjusting instrument with advanced features and 
functions.  Before assuming that unexpected behavior is caused by a system defect or breakdown, the 
operator should use this manual and other training facilities to become thoroughly familiar with  GPC1 
operation.  This troubleshooting guide is intended as an aid in identifying the cause of unexpected GPC1 
behavior and determining whether the behavior is due to normal operation or an internal or external 
problem. 

Identify the symptom or unexpected behavior you are observing from the 

SYMPTOMS

 listed in Table 3.   

PROBABLE

 

CAUSE

 is provided and a 

SOLUTION

 is proposed including references to manual 

sections that provide information that may be of assistance. 

Table 3.  

GPC1 Troubleshooting Checklist 

SYMPTOM 

PROBABLE CAUSE 

SOLUTION 

Test pressure continuously increases 
even with all valves closed. 

Leak in 

INLET valve

 or leak in 

PDVV 

(+) valve(s).

 

Isolate leak to 

INLET valve

 or 

PDVV (+) 

valve

(s) by checking whether pressure 

continues to increase when PDVV is at end 
of stoke.  Replace or repair valve(s) if qualified 
to do so. 3.1, 3.2.3. 

Test pressure continuously decreases 
even with all valves closed. 

Leak in 

OUTLET valve

, leak in 

PDVV (-) valve(s

), leak in test 

system connected to GPC1 or 
internal leak in GPC1 high pressure 
gas system. 

Identify and correct leak in test system if 
present.  Disconnect from test system and 
plug GPC1 

TEST ports 

to test GPC1 alone.  

Isolate leak to 

OUTLET valve

 or 

PDVV (-) 

valve(s)

 by checking if pressure continues to 

decrease when PDVV is in minimum end of 
stroke position.  Replace or repair valve(s) if 
qualified to do so. 3.1, 3.2.3. 

Test pressure takes too long to 
stabilize or will never stabilize. 

You are observing normal evolution 
of pressure in an uncontrolled static 
volume. 

Reduce resolution of pressure reading to 
appropriate level.  Reduce test volume. 

 

Increase stability of test tubings and vessels.  
Wait longer for stability.  Overshoot test point 
and return to reduce stabilization time. 

Test pressure takes too long to 
stabilize or will never stabilize. 

Leak in GPC1 or test system 
connected to GPC1. 

See troubleshooting tips “Test pressure 
continuously increases…” and “Test pressure 
continuously decreases…” above. 

Test pressure takes too long to 
stabilize or will never stabilize. 

GPC1 and/or test system tubing is 
contaminated with liquids that are 
“plugging” tubes. 

Thoroughly purge complete system. 
3.2.5 and 4.2 

PDVV will not increase pressure. 

PDVV is at maximum end of stroke 
position, PDVV drive air supply 
pressure is not high enough, or 
PDVV drive air supply is plugged. 

Verify PDVV piston position and readjust if 
necessary.  Use 

INLET valve

 to increase 

pressure.  Increase PDVV drive air supply 
pressure.  Clean filter. 2.3.2, 3.1, 3.2.3 

PDVV will not decrease pressure. 

PDVV  is  at  minimum  end  of  stroke 
position. 

Verify PDVV piston position and readjust if 
necessary.  Use 

OUTLET valve

 to decrease 

pressure. 2.3.2, 3.1, 3.2.3 

Opening 

INLET valve

 does not 

increase pressure. 

TEST INLET SET regulator

 is set 

too low and/or external pressure 
supply to low. 

Check value of supply pressure on 

TEST 

INLET gauge

.  Adjust 

INLET regulator

increase supply pressure if necessary. 2.3.2, 
3.2.1, Table 2 

Содержание GPC1-10000

Страница 1: ...2007 DH Instruments a Fluke Company GPC1 10000 GPC1 16000 Gas Pressure Controller Operation and Maintenance Manual...

Страница 2: ...and quality of its published materials however no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from...

Страница 3: ...ECT PNEUMATIC POWER DRIVE AIR 7 2 3 3 MAKE HIGH PRESSURE GAS SUPPLY CONNECTION 8 2 3 4 MAKE HIGH PRESSURE GAS INTERCONNECTIONS 8 2 3 4 1 CONNECTING TO A DHI RPM OR PG7202 USING THE STANDARD GPC1 INTER...

Страница 4: ...LET VALVE OPERATION 20 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALVE OPERATION 20 3 2 4 CONNECTING A DEVICE UNDER TEST DUT 21 3 2 5 PURGING LIQUIDS FROM THE DUT SYSTEM UNDER TEST 22 3 3 TYPICAL OPERAT...

Страница 5: ...orize d Service Providers 30 F FI IG GU UR RE ES S Figure 1 Front Panel View 3 Figure 2 Front and Side Views with Dimensions 3 Figure 3 System Schematic 4 Figure 4 Connecting GPC1 to PG7202 or RPM 9 F...

Страница 6: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Страница 7: ...and functions FOR THOSE OF YOU WHO DON T READ MANUALS GO DIRECTLY TO SECTION 2 3 TO SET UP YOUR GPC1 THEN GO TO SECTIONS 2 4 5 and 3 1 THIS WILL GET YOU RUNNING QUICKLY WITH MINIMAL RISK OF CAUSING DA...

Страница 8: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...

Страница 9: ...eference pressure monitor to configure a transfer standard based calibration system GPC1 requires an external pressure supply equal to the maximum pressure to be adjusted This is usually supplied by a...

Страница 10: ...a 120 psi Operating Temperature Range 10 to 35 C Weight 24 kg 53 lb Dimensions 21 5 cm H x 30 cm W x 53 5 cm D 8 5 in x 11 8 in x 21 1 in Pressure Range GPC1 16000 0 to 110 MPa 16 000 psi GPC1 10000 0...

Страница 11: ...and adjust pressure Figure 1 Front Panel View 1 3 2 OVERALL DIMENSIONS Figure 2 Front and Side Views with Dimensions 1 PDVV drive air test gas pressure gauge 2 PDVV piston position indicator 3 Test in...

Страница 12: ...EST INLET SET regulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13...

Страница 13: ...GPC1 Parts List DESCRIPTION PART GPC1 16000 or GCP1 16000 Hydrocarbon free version or GPC1 10000 or GPC1 10000 Hydrocarbon free version 401800 or 401801 402200 or 402201 ACCESSORIES INCLUDING 1 Mat T...

Страница 14: ...2 3 4 3 Unused TEST ports must be plugged for operation using the DH 500 plugs supplied in the GPC1 interconnections kit If GPC1 is to be used with a DHI RPM or PG7202 reference device the hardware n...

Страница 15: ...re summarized in Table 2 Since the flow requirements are very low and cleanliness of the gas is important the supply is usually regulated Nitrogen or instrument grade air from a compressed air bottle...

Страница 16: ...all connections have been properly tightened see Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 DH500 F fittings are d...

Страница 17: ...RE INTERCONNECTIONS GPC1 is delivered with an interconnections kit see Section 2 1 2 Table 1 This kit is designed to provide a standard connection to a DHI RPM or PG7202 see Section 2 3 4 1 It is conf...

Страница 18: ...return from the system to which it is connected For this purpose GPC1 is equipped with an inclined manifold that serves as a liquid sump see Figure 5 A DRAIN port is provided at the low point so that...

Страница 19: ...directly with clean gas from the SUPPLY port and exhausted away from the port Do not break the TEST1 port connection under pressure If unclean gas is present in GPC1 flowing out the TEST1 port will co...

Страница 20: ...e Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 2 4 2 ADJUST INLET PRESSURE This section assumes that the GPC1 system...

Страница 21: ...ference and or device under test should begin to increase If the pressure does not increase possible explanations are The TEST INLET SET regulator has not been properly adjusted see Section 2 4 2 Ther...

Страница 22: ...3 Drain purge the GPC1 sump regularly to remove any contaminants that may have entered the system from the supply gas or devices connected to GPC1 see Section 3 2 5 Put the PDVV plunger in the START p...

Страница 23: ...high pressure gas circuit Of course GPC1 will remain hydrocarbon free only so long as hydrocarbons are not introduced from an external source To maintain a hydrocarbon free GPC1 only connect devices...

Страница 24: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...

Страница 25: ...ose and counter clockwise CCW to open A WHITE dot on the handle body indicates the valve s current open close position The TEST INLET SET regulator 5 and TEST INLET gauge 4 are used to set the supply...

Страница 26: ...gulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13 Fast and slow P...

Страница 27: ...void accidental overpressure The PDVV can then be used for the final pressure adjustment at the maximum pressure see Section 3 2 3 With the INLET valve 5 closed use the INLET SET regulator 4 and the I...

Страница 28: ...n the system connected to GPC1 Use caution in opening the TEST INLET valve and always check the TEST INLET pressure gauge before doing so see Section 2 4 2 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALV...

Страница 29: ...vented The plunger is returned to minimum end of stroke position by a spring The PDVV plunger is usually set to the desired position at the start of a calibration or test The PDVV valves generate pres...

Страница 30: ...is as follows Connect the DUT to the appropriate GPC1 TEST port or to a test port on the external system to which GPC1 is connected see Sections 3 2 4 2 3 4 3 Set the TEST INLET SET regulator see Sect...

Страница 31: ...ntenance and repair services for GPC1 are offered by DHI Authorized Service Providers see Section 6 2 Table 4 GPC1 is a sophisticated pressure setting and adjusting instrument with advanced features a...

Страница 32: ...or liquid vapor Repeat steps through until there is no evidence of liquid in the gas that vents through the DRAIN port Figure 9 Drain Port Location 4 3 CLEANING REPLACING PDVV DRIVE AIR FILTER ELEMENT...

Страница 33: ...esent The filter may be replaced completely cleaned by back flushing or disassembled and the filter element cleaned or replaced To disassemble the filter body open the body by unscrewing the filter bo...

Страница 34: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 26 N NO OT TE ES S...

Страница 35: ...alone Isolate leak to OUTLET valve or PDVV valve s by checking if pressure continues to decrease when PDVV is in minimum end of stroke position Replace or repair valve s if qualified to do so 3 1 3 2...

Страница 36: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 28 N NO OT TE ES S...

Страница 37: ...rive pressure to the PDVV actuator causing the PDVV plunger to move forward compressing the high pressure gas and increasing the test pressure Used to increase pressure and for fine pressure adjustmen...

Страница 38: ...ue or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warr...

Отзывы: