background image

 

3.  OPERATION 

 

Page 17 

© 2007 DH Instruments, a Fluke Company 

3

3

.

.

 

 

O

O

P

P

E

E

R

R

A

A

T

T

I

I

O

O

N

N

 

 

3.1

 

OPERATING PRINCIPLE 

Numerical references in this section refer to Section 3.1, Figure 6. 

GPC1 is a self-contained system designed to set and adjust gas pressure into static pressure test and 
calibration systems.  GPC1-16000 operates from atmosphere  (zero gauge)  to  110 MPa  (16 000 psi).  
GPC1-10000 operates to 70 MPa (10 000 psi).  GPC1 combines the capability to execute large pressure 
changes very quickly with very fine pressure adjustment. 

GPC1 uses two different techniques to set and adjust pressure.  These are rough pressure control using 
the TEST INLET valve (12) and TEST OUTLET valve (15) and fine pressure control using the PDVV (+) 
(13) and (-) valves (14). 

The first means of generating and adjusting pressure uses the external gas pressure supply (1) with the 
TEST INLET SET regulator (5), TEST INLET valve (12) and TEST OUTLET valve (15).  This combination 
is used for filling the system under test, making large pressure changes and for rough pressure control.  
The valves are Belleville spring loaded, half-turn needle valves.  Their operation is highly progressive 
over a half-turn with mechanical stops at each end so they cannot be over tightened.  The regulator is self 
venting and its output is indicated by the TEST INLET gauge (4).  The valve is turned clockwise (CW) to 
close and counter-clockwise (CCW) to open.  A WHITE dot on the handle body indicates the valve’s 
current open/close position. The TEST INLET SET regulator (5) and TEST INLET gauge (4) are used to 
set the supply pressure to the TEST INLET valve (12).  Opening the INLET valve (12) allows the supply 
pressure to enter GPC1 and reach the TEST ports (6,7,8).  Opening the OUTLET valve (15) exhausts 
gas, reducing pressure in the system. 

The second means of generating and adjusting pressure is the Pneumatically Driven Variable Volume 
(PDVV) (16) combined with the PDVV (+) (13) and (-) valves (14).  This combination is used for smaller 
pressure changes and fine pressure control.  The PDVV is a pneumatically actuated variable volume.  A 
piston or plunger in a cylinder is exposed to the high pressure gas to be adjusted. 

 

 

The other end of the plunger is connected to a dome loaded pneumatic actuator.  Changing the 
pneumatic pressure on the dome loaded actuator causes the plunger to move, increasing or decreasing 
the high gas pressure.  A spring returns the plunger to its minimum stroke position when there is no 
pressure on the dome.  A mechanical system tracks movement of the plunger and an indicator (18) 
displays the plunger position on the front panel.  The PDVV (+) (13) and (-) valves (14) are momentary, 
poppet valves that open when pressed.  The (+) valves (13) admit drive air pressure to the PDVV 
actuator causing the PDVV piston to move forward, compressing the high pressure gas and increasing 
the pressure.  The (-) valves (14) have the opposite effect, causing pressure to decrease when they are 
operated. 

Содержание GPC1-10000

Страница 1: ...2007 DH Instruments a Fluke Company GPC1 10000 GPC1 16000 Gas Pressure Controller Operation and Maintenance Manual...

Страница 2: ...and quality of its published materials however no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from...

Страница 3: ...ECT PNEUMATIC POWER DRIVE AIR 7 2 3 3 MAKE HIGH PRESSURE GAS SUPPLY CONNECTION 8 2 3 4 MAKE HIGH PRESSURE GAS INTERCONNECTIONS 8 2 3 4 1 CONNECTING TO A DHI RPM OR PG7202 USING THE STANDARD GPC1 INTER...

Страница 4: ...LET VALVE OPERATION 20 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALVE OPERATION 20 3 2 4 CONNECTING A DEVICE UNDER TEST DUT 21 3 2 5 PURGING LIQUIDS FROM THE DUT SYSTEM UNDER TEST 22 3 3 TYPICAL OPERAT...

Страница 5: ...orize d Service Providers 30 F FI IG GU UR RE ES S Figure 1 Front Panel View 3 Figure 2 Front and Side Views with Dimensions 3 Figure 3 System Schematic 4 Figure 4 Connecting GPC1 to PG7202 or RPM 9 F...

Страница 6: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Страница 7: ...and functions FOR THOSE OF YOU WHO DON T READ MANUALS GO DIRECTLY TO SECTION 2 3 TO SET UP YOUR GPC1 THEN GO TO SECTIONS 2 4 5 and 3 1 THIS WILL GET YOU RUNNING QUICKLY WITH MINIMAL RISK OF CAUSING DA...

Страница 8: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...

Страница 9: ...eference pressure monitor to configure a transfer standard based calibration system GPC1 requires an external pressure supply equal to the maximum pressure to be adjusted This is usually supplied by a...

Страница 10: ...a 120 psi Operating Temperature Range 10 to 35 C Weight 24 kg 53 lb Dimensions 21 5 cm H x 30 cm W x 53 5 cm D 8 5 in x 11 8 in x 21 1 in Pressure Range GPC1 16000 0 to 110 MPa 16 000 psi GPC1 10000 0...

Страница 11: ...and adjust pressure Figure 1 Front Panel View 1 3 2 OVERALL DIMENSIONS Figure 2 Front and Side Views with Dimensions 1 PDVV drive air test gas pressure gauge 2 PDVV piston position indicator 3 Test in...

Страница 12: ...EST INLET SET regulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13...

Страница 13: ...GPC1 Parts List DESCRIPTION PART GPC1 16000 or GCP1 16000 Hydrocarbon free version or GPC1 10000 or GPC1 10000 Hydrocarbon free version 401800 or 401801 402200 or 402201 ACCESSORIES INCLUDING 1 Mat T...

Страница 14: ...2 3 4 3 Unused TEST ports must be plugged for operation using the DH 500 plugs supplied in the GPC1 interconnections kit If GPC1 is to be used with a DHI RPM or PG7202 reference device the hardware n...

Страница 15: ...re summarized in Table 2 Since the flow requirements are very low and cleanliness of the gas is important the supply is usually regulated Nitrogen or instrument grade air from a compressed air bottle...

Страница 16: ...all connections have been properly tightened see Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 DH500 F fittings are d...

Страница 17: ...RE INTERCONNECTIONS GPC1 is delivered with an interconnections kit see Section 2 1 2 Table 1 This kit is designed to provide a standard connection to a DHI RPM or PG7202 see Section 2 3 4 1 It is conf...

Страница 18: ...return from the system to which it is connected For this purpose GPC1 is equipped with an inclined manifold that serves as a liquid sump see Figure 5 A DRAIN port is provided at the low point so that...

Страница 19: ...directly with clean gas from the SUPPLY port and exhausted away from the port Do not break the TEST1 port connection under pressure If unclean gas is present in GPC1 flowing out the TEST1 port will co...

Страница 20: ...e Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 2 4 2 ADJUST INLET PRESSURE This section assumes that the GPC1 system...

Страница 21: ...ference and or device under test should begin to increase If the pressure does not increase possible explanations are The TEST INLET SET regulator has not been properly adjusted see Section 2 4 2 Ther...

Страница 22: ...3 Drain purge the GPC1 sump regularly to remove any contaminants that may have entered the system from the supply gas or devices connected to GPC1 see Section 3 2 5 Put the PDVV plunger in the START p...

Страница 23: ...high pressure gas circuit Of course GPC1 will remain hydrocarbon free only so long as hydrocarbons are not introduced from an external source To maintain a hydrocarbon free GPC1 only connect devices...

Страница 24: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...

Страница 25: ...ose and counter clockwise CCW to open A WHITE dot on the handle body indicates the valve s current open close position The TEST INLET SET regulator 5 and TEST INLET gauge 4 are used to set the supply...

Страница 26: ...gulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13 Fast and slow P...

Страница 27: ...void accidental overpressure The PDVV can then be used for the final pressure adjustment at the maximum pressure see Section 3 2 3 With the INLET valve 5 closed use the INLET SET regulator 4 and the I...

Страница 28: ...n the system connected to GPC1 Use caution in opening the TEST INLET valve and always check the TEST INLET pressure gauge before doing so see Section 2 4 2 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALV...

Страница 29: ...vented The plunger is returned to minimum end of stroke position by a spring The PDVV plunger is usually set to the desired position at the start of a calibration or test The PDVV valves generate pres...

Страница 30: ...is as follows Connect the DUT to the appropriate GPC1 TEST port or to a test port on the external system to which GPC1 is connected see Sections 3 2 4 2 3 4 3 Set the TEST INLET SET regulator see Sect...

Страница 31: ...ntenance and repair services for GPC1 are offered by DHI Authorized Service Providers see Section 6 2 Table 4 GPC1 is a sophisticated pressure setting and adjusting instrument with advanced features a...

Страница 32: ...or liquid vapor Repeat steps through until there is no evidence of liquid in the gas that vents through the DRAIN port Figure 9 Drain Port Location 4 3 CLEANING REPLACING PDVV DRIVE AIR FILTER ELEMENT...

Страница 33: ...esent The filter may be replaced completely cleaned by back flushing or disassembled and the filter element cleaned or replaced To disassemble the filter body open the body by unscrewing the filter bo...

Страница 34: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 26 N NO OT TE ES S...

Страница 35: ...alone Isolate leak to OUTLET valve or PDVV valve s by checking if pressure continues to decrease when PDVV is in minimum end of stroke position Replace or repair valve s if qualified to do so 3 1 3 2...

Страница 36: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 28 N NO OT TE ES S...

Страница 37: ...rive pressure to the PDVV actuator causing the PDVV plunger to move forward compressing the high pressure gas and increasing the test pressure Used to increase pressure and for fine pressure adjustmen...

Страница 38: ...ue or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warr...

Отзывы: