background image

GPC1-10000 / GPC1-16000

 OPERATION AND MAINTENANCE MANUAL 

© 2007 DH Instruments, a Fluke Company 

Page 20

 

3.2.2

 

ROUGH PRESSURE GENERATION/CONTROL, INLET 

AND OUTLET VALVE OPERATION 

Numerical references in this section refer to Section 3.2,  Figure 7.  For information on the 
GPC1 operating principle, see Section 3.1. 

The 

INLET valve 

(5) and 

OUTLET valve 

(8) are used to execute large pressure changes in the 

test system and for rough pressure control.  The 

INLET valve

 is used to increase pressure and 

the 

OUTLET valve

, decreases pressure and vents the system to atmosphere, setting zero gauge 

pressure.  If finer pressure control, use the 

PDVV valves 

(see Section 3.2.3). 

To increase pressure in the GPC1 system, open the 

INLET valve 

(5) slowly.  Use the 

progressive nature of the valve to control the rate of pressure increase.  Close the valve 
when the desired pressure has been reached.  When a gas booster is used to supply high 
pressure, the booster may not be able to keep up with the flow rate needed to set the desired 
pressure.  In this case, with the 

INLET valve

 open, wait for the booster to reach the desired 

pressure. 

To decrease pressure in the GPC1 system, open 

OUTLET valve 

(8) slowly.  Open the valve 

fully to vent the system to atmosphere (set zero gauge). 

 

 

The TEST INLET valve connects the pressure supply to the test system.  When the TEST 
INLET valve is opened, high pressure can be generated rapidly in the system connected to 
GPC1. Use caution in opening the TEST INLET valve and always check the TEST INLET 
pressure gauge before doing so (see Section 2.4.2). 

 

3.2.3

 

FINE PRESSURE ADJUSTMENT, PDVV (+) AND (-) 

VALVE OPERATION 

Numerical references in this section refer to Section 3.2,  Figure 7 except where 
specified otherwise.  For information on the GPC1 operating principle, see Section 3.1. 

Use the valves labeled 

 

and 

 to release gas from the PDVV actuator, causing the PDVV 

plunger to back off and pressure to decrease.  Use the valves labeled 

 

are

 

 to admit gas 

to the PDVV actuator, causing the PDVV plunger to move forward and pressure to increase.  
The 

 and 

 

valves are for high speed PDVV operation

.  

The 

 

and 

 valves are for slow speed 

PDVV operation. 

The 

PDVV (+) 

(6) and 

(-) valves 

(7) are utilized to make small pressure changes and for fine 

pressure control.  Brief momentary action on the 

 

and 

 valves is used to 

bump 

or 

jog

 

pressure in very small amounts around a pressure point. 

If the PDVV plunger reaches end of stroke during a calibration or test, use the 

INLET valve 

(5) 

and/or 

OUTLET valve 

(8) to increase or decrease the pressure, as needed (see Section 3.2.2). 

The actual rate of pressure change caused by the 

PDVV (+) valves

 and 

PDVV (-) valves

 is 

dependent on the test volume that is connected to GPC1 and the current pressure value.  
Increasing the test volume lowers the pressure change rate and pressure step size.  
Increasing the volume increases the rate and the pressure step size.  The maximum 
pressure that can be generated by the 

PDVV (+) valves

 is dependent on the PDVV drive air 

supply (see Section 2.3.2, Table 2). 

 

Содержание GPC1-10000

Страница 1: ...2007 DH Instruments a Fluke Company GPC1 10000 GPC1 16000 Gas Pressure Controller Operation and Maintenance Manual...

Страница 2: ...and quality of its published materials however no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from...

Страница 3: ...ECT PNEUMATIC POWER DRIVE AIR 7 2 3 3 MAKE HIGH PRESSURE GAS SUPPLY CONNECTION 8 2 3 4 MAKE HIGH PRESSURE GAS INTERCONNECTIONS 8 2 3 4 1 CONNECTING TO A DHI RPM OR PG7202 USING THE STANDARD GPC1 INTER...

Страница 4: ...LET VALVE OPERATION 20 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALVE OPERATION 20 3 2 4 CONNECTING A DEVICE UNDER TEST DUT 21 3 2 5 PURGING LIQUIDS FROM THE DUT SYSTEM UNDER TEST 22 3 3 TYPICAL OPERAT...

Страница 5: ...orize d Service Providers 30 F FI IG GU UR RE ES S Figure 1 Front Panel View 3 Figure 2 Front and Side Views with Dimensions 3 Figure 3 System Schematic 4 Figure 4 Connecting GPC1 to PG7202 or RPM 9 F...

Страница 6: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Страница 7: ...and functions FOR THOSE OF YOU WHO DON T READ MANUALS GO DIRECTLY TO SECTION 2 3 TO SET UP YOUR GPC1 THEN GO TO SECTIONS 2 4 5 and 3 1 THIS WILL GET YOU RUNNING QUICKLY WITH MINIMAL RISK OF CAUSING DA...

Страница 8: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...

Страница 9: ...eference pressure monitor to configure a transfer standard based calibration system GPC1 requires an external pressure supply equal to the maximum pressure to be adjusted This is usually supplied by a...

Страница 10: ...a 120 psi Operating Temperature Range 10 to 35 C Weight 24 kg 53 lb Dimensions 21 5 cm H x 30 cm W x 53 5 cm D 8 5 in x 11 8 in x 21 1 in Pressure Range GPC1 16000 0 to 110 MPa 16 000 psi GPC1 10000 0...

Страница 11: ...and adjust pressure Figure 1 Front Panel View 1 3 2 OVERALL DIMENSIONS Figure 2 Front and Side Views with Dimensions 1 PDVV drive air test gas pressure gauge 2 PDVV piston position indicator 3 Test in...

Страница 12: ...EST INLET SET regulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13...

Страница 13: ...GPC1 Parts List DESCRIPTION PART GPC1 16000 or GCP1 16000 Hydrocarbon free version or GPC1 10000 or GPC1 10000 Hydrocarbon free version 401800 or 401801 402200 or 402201 ACCESSORIES INCLUDING 1 Mat T...

Страница 14: ...2 3 4 3 Unused TEST ports must be plugged for operation using the DH 500 plugs supplied in the GPC1 interconnections kit If GPC1 is to be used with a DHI RPM or PG7202 reference device the hardware n...

Страница 15: ...re summarized in Table 2 Since the flow requirements are very low and cleanliness of the gas is important the supply is usually regulated Nitrogen or instrument grade air from a compressed air bottle...

Страница 16: ...all connections have been properly tightened see Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 DH500 F fittings are d...

Страница 17: ...RE INTERCONNECTIONS GPC1 is delivered with an interconnections kit see Section 2 1 2 Table 1 This kit is designed to provide a standard connection to a DHI RPM or PG7202 see Section 2 3 4 1 It is conf...

Страница 18: ...return from the system to which it is connected For this purpose GPC1 is equipped with an inclined manifold that serves as a liquid sump see Figure 5 A DRAIN port is provided at the low point so that...

Страница 19: ...directly with clean gas from the SUPPLY port and exhausted away from the port Do not break the TEST1 port connection under pressure If unclean gas is present in GPC1 flowing out the TEST1 port will co...

Страница 20: ...e Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 2 4 2 ADJUST INLET PRESSURE This section assumes that the GPC1 system...

Страница 21: ...ference and or device under test should begin to increase If the pressure does not increase possible explanations are The TEST INLET SET regulator has not been properly adjusted see Section 2 4 2 Ther...

Страница 22: ...3 Drain purge the GPC1 sump regularly to remove any contaminants that may have entered the system from the supply gas or devices connected to GPC1 see Section 3 2 5 Put the PDVV plunger in the START p...

Страница 23: ...high pressure gas circuit Of course GPC1 will remain hydrocarbon free only so long as hydrocarbons are not introduced from an external source To maintain a hydrocarbon free GPC1 only connect devices...

Страница 24: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...

Страница 25: ...ose and counter clockwise CCW to open A WHITE dot on the handle body indicates the valve s current open close position The TEST INLET SET regulator 5 and TEST INLET gauge 4 are used to set the supply...

Страница 26: ...gulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13 Fast and slow P...

Страница 27: ...void accidental overpressure The PDVV can then be used for the final pressure adjustment at the maximum pressure see Section 3 2 3 With the INLET valve 5 closed use the INLET SET regulator 4 and the I...

Страница 28: ...n the system connected to GPC1 Use caution in opening the TEST INLET valve and always check the TEST INLET pressure gauge before doing so see Section 2 4 2 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALV...

Страница 29: ...vented The plunger is returned to minimum end of stroke position by a spring The PDVV plunger is usually set to the desired position at the start of a calibration or test The PDVV valves generate pres...

Страница 30: ...is as follows Connect the DUT to the appropriate GPC1 TEST port or to a test port on the external system to which GPC1 is connected see Sections 3 2 4 2 3 4 3 Set the TEST INLET SET regulator see Sect...

Страница 31: ...ntenance and repair services for GPC1 are offered by DHI Authorized Service Providers see Section 6 2 Table 4 GPC1 is a sophisticated pressure setting and adjusting instrument with advanced features a...

Страница 32: ...or liquid vapor Repeat steps through until there is no evidence of liquid in the gas that vents through the DRAIN port Figure 9 Drain Port Location 4 3 CLEANING REPLACING PDVV DRIVE AIR FILTER ELEMENT...

Страница 33: ...esent The filter may be replaced completely cleaned by back flushing or disassembled and the filter element cleaned or replaced To disassemble the filter body open the body by unscrewing the filter bo...

Страница 34: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 26 N NO OT TE ES S...

Страница 35: ...alone Isolate leak to OUTLET valve or PDVV valve s by checking if pressure continues to decrease when PDVV is in minimum end of stroke position Replace or repair valve s if qualified to do so 3 1 3 2...

Страница 36: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 28 N NO OT TE ES S...

Страница 37: ...rive pressure to the PDVV actuator causing the PDVV plunger to move forward compressing the high pressure gas and increasing the test pressure Used to increase pressure and for fine pressure adjustmen...

Страница 38: ...ue or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warr...

Отзывы: