background image

 

2. 5BINSTALLATION 

 

Page 7 

© 2007 DH Instruments, a Fluke Company

 

2.3

 

INITIAL SETUP 

 

  

Before setting up GPC1, see Section 2.2 for general information on site requirements. 

 

To prepare a GPC1 for check out and operation: 

  Set up GPC1 (see Section 2.3.1). 

  Connect drive air (see Section 2.3.2). 

  Connect the high pressure supply (see Section 2.3.3) 

  Make the system high pressure gas interconnections (see Section 2.3.4). 

2.3.1

 

SET UP GPC1 

To set up GPC1 proceed as follows: 

  Place GPC1 on the site table in the proper orientation with the front panel controls 

conveniently accessible. 

  Install the mat delivered with the GPC1 accessories on the top, front surface of the 

GPC1. 

2.3.2

 

CONNECT PNEUMATIC POWER (DRIVE AIR) 

GPC1 requires pneumatic power to drive its pneumatically driven variable volume (PDVV) 
(see Figure 3). 

The drive air requirements are summarized in Table 2.  Since the flow requirements are very 
low and cleanliness of the gas is important, the supply is usually regulated Nitrogen or 
instrument grade air from a compressed air bottle. 

Connect the pneumatic pressure source to the 1/4 in. NPT F connection labeled 

PDVV

 

DRIVE 

on the right side of GPC1 housing.  Use Teflon

 tape or another thread sealant to 

minimize leakage. 

 

 

GPC1 must be supplied with NON-LUBRICATED drive air.  The internal components are 
permanently lubricated.  The oil in lubricated air can contaminate the small diameter 
tubing inside GPC1 and lead to erratic behavior requiring difficult and costly cleaning. 

 

Table 2.  

Pneumatic Power (Drive Air) Requirements 

INTERNAL 

COMPONENT 

DRIVE AIR  

PRESSURE NEEDED 

MINIMUM 

FLOW 

LUBRICATION 

REQUIRED 

CLEANLINESS

CRITICAL 

PDVV 

For 5 MPa (750 psi):   
 

Supply 300 kPa  (45 psi) 

For 50 MPa (7 500 psi):  
 

Supply 600 kPa (87 psi) 

For 70 MPa (10 000 psi): 
 

Supply 700 kPa (100 psi) 

For 110 MPa (16 000 psi): 
  

Supply 850 kPa (120 psi) 

200 sccm (< 0.01 scfm) 

* This is very low flow. 

No Yes 

Содержание GPC1-10000

Страница 1: ...2007 DH Instruments a Fluke Company GPC1 10000 GPC1 16000 Gas Pressure Controller Operation and Maintenance Manual...

Страница 2: ...and quality of its published materials however no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from...

Страница 3: ...ECT PNEUMATIC POWER DRIVE AIR 7 2 3 3 MAKE HIGH PRESSURE GAS SUPPLY CONNECTION 8 2 3 4 MAKE HIGH PRESSURE GAS INTERCONNECTIONS 8 2 3 4 1 CONNECTING TO A DHI RPM OR PG7202 USING THE STANDARD GPC1 INTER...

Страница 4: ...LET VALVE OPERATION 20 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALVE OPERATION 20 3 2 4 CONNECTING A DEVICE UNDER TEST DUT 21 3 2 5 PURGING LIQUIDS FROM THE DUT SYSTEM UNDER TEST 22 3 3 TYPICAL OPERAT...

Страница 5: ...orize d Service Providers 30 F FI IG GU UR RE ES S Figure 1 Front Panel View 3 Figure 2 Front and Side Views with Dimensions 3 Figure 3 System Schematic 4 Figure 4 Connecting GPC1 to PG7202 or RPM 9 F...

Страница 6: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Страница 7: ...and functions FOR THOSE OF YOU WHO DON T READ MANUALS GO DIRECTLY TO SECTION 2 3 TO SET UP YOUR GPC1 THEN GO TO SECTIONS 2 4 5 and 3 1 THIS WILL GET YOU RUNNING QUICKLY WITH MINIMAL RISK OF CAUSING DA...

Страница 8: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...

Страница 9: ...eference pressure monitor to configure a transfer standard based calibration system GPC1 requires an external pressure supply equal to the maximum pressure to be adjusted This is usually supplied by a...

Страница 10: ...a 120 psi Operating Temperature Range 10 to 35 C Weight 24 kg 53 lb Dimensions 21 5 cm H x 30 cm W x 53 5 cm D 8 5 in x 11 8 in x 21 1 in Pressure Range GPC1 16000 0 to 110 MPa 16 000 psi GPC1 10000 0...

Страница 11: ...and adjust pressure Figure 1 Front Panel View 1 3 2 OVERALL DIMENSIONS Figure 2 Front and Side Views with Dimensions 1 PDVV drive air test gas pressure gauge 2 PDVV piston position indicator 3 Test in...

Страница 12: ...EST INLET SET regulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13...

Страница 13: ...GPC1 Parts List DESCRIPTION PART GPC1 16000 or GCP1 16000 Hydrocarbon free version or GPC1 10000 or GPC1 10000 Hydrocarbon free version 401800 or 401801 402200 or 402201 ACCESSORIES INCLUDING 1 Mat T...

Страница 14: ...2 3 4 3 Unused TEST ports must be plugged for operation using the DH 500 plugs supplied in the GPC1 interconnections kit If GPC1 is to be used with a DHI RPM or PG7202 reference device the hardware n...

Страница 15: ...re summarized in Table 2 Since the flow requirements are very low and cleanliness of the gas is important the supply is usually regulated Nitrogen or instrument grade air from a compressed air bottle...

Страница 16: ...all connections have been properly tightened see Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 DH500 F fittings are d...

Страница 17: ...RE INTERCONNECTIONS GPC1 is delivered with an interconnections kit see Section 2 1 2 Table 1 This kit is designed to provide a standard connection to a DHI RPM or PG7202 see Section 2 3 4 1 It is conf...

Страница 18: ...return from the system to which it is connected For this purpose GPC1 is equipped with an inclined manifold that serves as a liquid sump see Figure 5 A DRAIN port is provided at the low point so that...

Страница 19: ...directly with clean gas from the SUPPLY port and exhausted away from the port Do not break the TEST1 port connection under pressure If unclean gas is present in GPC1 flowing out the TEST1 port will co...

Страница 20: ...e Section 2 4 5 Ensure that the GPC1 TEST INLET SET regulator is fully backed off and the INLET valve is closed see Section 2 4 1 2 4 2 ADJUST INLET PRESSURE This section assumes that the GPC1 system...

Страница 21: ...ference and or device under test should begin to increase If the pressure does not increase possible explanations are The TEST INLET SET regulator has not been properly adjusted see Section 2 4 2 Ther...

Страница 22: ...3 Drain purge the GPC1 sump regularly to remove any contaminants that may have entered the system from the supply gas or devices connected to GPC1 see Section 3 2 5 Put the PDVV plunger in the START p...

Страница 23: ...high pressure gas circuit Of course GPC1 will remain hydrocarbon free only so long as hydrocarbons are not introduced from an external source To maintain a hydrocarbon free GPC1 only connect devices...

Страница 24: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...

Страница 25: ...ose and counter clockwise CCW to open A WHITE dot on the handle body indicates the valve s current open close position The TEST INLET SET regulator 5 and TEST INLET gauge 4 are used to set the supply...

Страница 26: ...gulator 6 TEST1 port rear top 7 TEST2 port left lateral 8 TEST3 port front top 9 DRAIN port right lateral 10 Test liquid collection sump 11 Sump filter not shown 12 TEST INLET valve 13 Fast and slow P...

Страница 27: ...void accidental overpressure The PDVV can then be used for the final pressure adjustment at the maximum pressure see Section 3 2 3 With the INLET valve 5 closed use the INLET SET regulator 4 and the I...

Страница 28: ...n the system connected to GPC1 Use caution in opening the TEST INLET valve and always check the TEST INLET pressure gauge before doing so see Section 2 4 2 3 2 3 FINE PRESSURE ADJUSTMENT PDVV AND VALV...

Страница 29: ...vented The plunger is returned to minimum end of stroke position by a spring The PDVV plunger is usually set to the desired position at the start of a calibration or test The PDVV valves generate pres...

Страница 30: ...is as follows Connect the DUT to the appropriate GPC1 TEST port or to a test port on the external system to which GPC1 is connected see Sections 3 2 4 2 3 4 3 Set the TEST INLET SET regulator see Sect...

Страница 31: ...ntenance and repair services for GPC1 are offered by DHI Authorized Service Providers see Section 6 2 Table 4 GPC1 is a sophisticated pressure setting and adjusting instrument with advanced features a...

Страница 32: ...or liquid vapor Repeat steps through until there is no evidence of liquid in the gas that vents through the DRAIN port Figure 9 Drain Port Location 4 3 CLEANING REPLACING PDVV DRIVE AIR FILTER ELEMENT...

Страница 33: ...esent The filter may be replaced completely cleaned by back flushing or disassembled and the filter element cleaned or replaced To disassemble the filter body open the body by unscrewing the filter bo...

Страница 34: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 26 N NO OT TE ES S...

Страница 35: ...alone Isolate leak to OUTLET valve or PDVV valve s by checking if pressure continues to decrease when PDVV is in minimum end of stroke position Replace or repair valve s if qualified to do so 3 1 3 2...

Страница 36: ...GPC1 10000 GPC1 16000 OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 28 N NO OT TE ES S...

Страница 37: ...rive pressure to the PDVV actuator causing the PDVV plunger to move forward compressing the high pressure gas and increasing the test pressure Used to increase pressure and for fine pressure adjustmen...

Страница 38: ...ue or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warr...

Отзывы: