![Dexcom G4 Platinum Professional Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/dexcom/g4-platinum-professional/g4-platinum-professional_user-manual_2490094054.webp)
Dexcom G4 PLATINUM Professional System User’s Guide
|
53
(1) Press the
UP
or
DOWN
button to select the alert level. The low
alert can be changed in steps of 5 mg/dL, and the high alert can
be changed in steps of 10 mg/dL.
(2) Press the
SELECT
button to confirm the alert level.
• You can also change the alert levels in the Alerts menu.
• The unit of measurement (mg/dL) is not adjustable.
• If you need to change the time, date or transmitter ID after
you complete the Setup Wizard see Section 5.2, The Settings
Menu.
The Setup Wizard is now complete. To start using the Dexcom G4
PLATINUM Professional CGM System you must insert a sensor (see
Chapter 6, Inserting a Sensor and Starting a Sensor Session).
5.2 THE SETTINGS MENU
The Settings menu lets you change the time, date or transmitter ID. The
Setup Wizard only works the first time you turn on the receiver, but you
can always use the Settings menu to make changes to the transmitter
ID, time and date, etc.
5.2.1 GETTING TO THE SETTINGS MENU
1. Press the
SELECT
button to turn on the receiver. The 3-hour trend
WARNING
Low/high glucose alerts and the 55 mg/dL low glucose
alarm are all disabled when using the Dexcom G4
PLATINUM Professional System in blinded mode and
will not alert you when the glucose is low or high. Check
blood glucose meter if there are any signs or symptoms
of low or high glucose.
Содержание G4 Platinum Professional
Страница 13: ...12 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1...
Страница 23: ...22 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 24: ...Dexcom G4 PLATINUM Professional System User s Guide 23 2 chapter two INDICATIONS FOR USE AND SAFETY STATEMENT...
Страница 25: ...24 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 2...
Страница 35: ...34 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...Dexcom G4 PLATINUM Professional System User s Guide 35 3 chapter three RISKS AND BENEFITS...
Страница 37: ...36 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1 3...
Страница 43: ...42 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 4...
Страница 51: ...50 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 5...
Страница 65: ...64 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 6...
Страница 81: ...80 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 7...
Страница 93: ...92 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10 8...
Страница 103: ...102 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 104: ...Dexcom G4 PLATINUM Professional System User s Guide 103 9 chapter nine ALERTS ALARM PROFILES...
Страница 105: ...104 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 9...
Страница 125: ...124 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 126: ...Dexcom G4 PLATINUM Professional System User s Guide 125 10 chapter ten EVENTS DEXCOM STUDIO SOFTWARE AND BLINDED MODE...
Страница 127: ...126 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10...
Страница 140: ...Dexcom G4 PLATINUM Professional System User s Guide 139 11 chapter eleven ENDING A SENSOR SESSION...
Страница 141: ...140 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 11...
Страница 147: ...146 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 149: ...148 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 12...
Страница 164: ...Dexcom G4 PLATINUM Professional System User s Guide 163 13 chapter thirteen TROUBLESHOOTING...
Страница 165: ...164 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 13...
Страница 182: ...Dexcom G4 PLATINUM Professional System User s Guide 181 14 chapter fourteen TECHNICAL INFORMATION...
Страница 183: ...182 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 14...
Страница 218: ...Dexcom G4 PLATINUM Professional System User s Guide 217 15 chapter fifteen USER ASSISTANCE...
Страница 219: ...218 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 15...
Страница 222: ...Dexcom G4 PLATINUM Professional System User s Guide 221 16 chapter sixteen WARRANTY...
Страница 223: ...222 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 16...
Страница 231: ...230 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 232: ...Dexcom G4 PLATINUM Professional System User s Guide 231 17 chapter seventeen TRAVEL INFORMATION...
Страница 233: ...232 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 17...
Страница 235: ...234 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 236: ...Dexcom G4 PLATINUM Professional System User s Guide 235 18 chapter eighteen APPENDIX...
Страница 237: ...236 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 18...
Страница 259: ...258 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...