![Dexcom G4 Platinum Professional Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/dexcom/g4-platinum-professional/g4-platinum-professional_user-manual_2490094038.webp)
Dexcom G4 PLATINUM Professional System User’s Guide
|
37
CHAPTER 3: RISKS AND BENEFITS
3.1 RISKS
There are some known risks with using real-time or blinded
CGM.
You will not get sensor glucose alerts when the alert function
is turned off, your transmitter and receiver are out of range,
or when your receiver is not showing sensor glucose
readings. You might not notice alerts if you are unable to
hear them or feel the vibration. Sometimes your sensor
glucose reading may be slightly different than your blood
glucose meter. For example, your blood glucose meter may
show a blood glucose value of 78 mg/dL, but your sensor
glucose reading may show as 82 mg/dL. If your low alert is
set at 80 mg/dL, your system will not alert for a low glucose
at this time because 82 mg/dL is still above your low alert
setting. In most cases, the sensor glucose readings will
move in the right direction and will alert you of a high or low
shortly after. If you do not get an alert for any reason, and
you do not take frequent blood glucose measurements with
your blood glucose meter, you might not be aware of low or
high blood glucose levels. If this happens, and your blood
glucose levels are very high or low, there is a remote chance
you might need medical help. There is a risk of disease
transmission when the device is used on multiple patients.
When the CGM system is used in blinded mode, the sensor
glucose readings are not displayed, only recorded for
review when downloaded later by the Dexcom STUDIO
software. Using the system in blinded mode might cause
you to expect glucose alerts, but you will not receive high
Содержание G4 Platinum Professional
Страница 13: ...12 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1...
Страница 23: ...22 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 24: ...Dexcom G4 PLATINUM Professional System User s Guide 23 2 chapter two INDICATIONS FOR USE AND SAFETY STATEMENT...
Страница 25: ...24 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 2...
Страница 35: ...34 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...Dexcom G4 PLATINUM Professional System User s Guide 35 3 chapter three RISKS AND BENEFITS...
Страница 37: ...36 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1 3...
Страница 43: ...42 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 4...
Страница 51: ...50 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 5...
Страница 65: ...64 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 6...
Страница 81: ...80 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 7...
Страница 93: ...92 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10 8...
Страница 103: ...102 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 104: ...Dexcom G4 PLATINUM Professional System User s Guide 103 9 chapter nine ALERTS ALARM PROFILES...
Страница 105: ...104 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 9...
Страница 125: ...124 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 126: ...Dexcom G4 PLATINUM Professional System User s Guide 125 10 chapter ten EVENTS DEXCOM STUDIO SOFTWARE AND BLINDED MODE...
Страница 127: ...126 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10...
Страница 140: ...Dexcom G4 PLATINUM Professional System User s Guide 139 11 chapter eleven ENDING A SENSOR SESSION...
Страница 141: ...140 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 11...
Страница 147: ...146 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 149: ...148 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 12...
Страница 164: ...Dexcom G4 PLATINUM Professional System User s Guide 163 13 chapter thirteen TROUBLESHOOTING...
Страница 165: ...164 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 13...
Страница 182: ...Dexcom G4 PLATINUM Professional System User s Guide 181 14 chapter fourteen TECHNICAL INFORMATION...
Страница 183: ...182 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 14...
Страница 218: ...Dexcom G4 PLATINUM Professional System User s Guide 217 15 chapter fifteen USER ASSISTANCE...
Страница 219: ...218 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 15...
Страница 222: ...Dexcom G4 PLATINUM Professional System User s Guide 221 16 chapter sixteen WARRANTY...
Страница 223: ...222 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 16...
Страница 231: ...230 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 232: ...Dexcom G4 PLATINUM Professional System User s Guide 231 17 chapter seventeen TRAVEL INFORMATION...
Страница 233: ...232 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 17...
Страница 235: ...234 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 236: ...Dexcom G4 PLATINUM Professional System User s Guide 235 18 chapter eighteen APPENDIX...
Страница 237: ...236 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 18...
Страница 259: ...258 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...