![Dexcom G4 Platinum Professional Скачать руководство пользователя страница 195](http://html1.mh-extra.com/html/dexcom/g4-platinum-professional/g4-platinum-professional_user-manual_2490094195.webp)
194
|
Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
Hypoglycemia Alert Rate
The Alert Rate shows how often the alert is right or wrong. The True
Alert Rate is the % of time the device alarmed when the blood glucose
level was at or below the alert setting within 15 minutes before or after
the device alarmed. The False Alert Rate is the % of time the device
alarmed when the blood glucose level was above the alert setting within
15 minutes before or after the device alarmed.
For example, if you set the Low Glucose Alert to 70 mg/dL (3.9 mmol/L)
and your alarm sounds, how often can you expect your blood sugar to
actually be low? If your alarm sounds, you can expect your blood sugar
to be below 70 mg/dL (3.9 mmol/L) approximately 79% of the time and
not be below 70 mg/dL (3.9 mmol/L) approximately 21% of the time
within the 15 minute period before or after your alarm sounds.
Hypoglycemia Detection Rate
The Detection Rate shows how often the device recognizes and alerts
you to an episode of hypoglycemia or how often it misses such an event.
The Hypoglycemia Detection Rate is the % of time the blood glucose
level was at or below the alert setting and device alarmed within 15
minutes before or after the blood glucose was at or below the alert
settings. The Hypoglycemia Missed Detection Rate is the % of time
the blood glucose was at or below the alert setting, but the device did
not alarm within 15 minutes before or after the blood glucose was at or
below the alert setting.
For example, if you set the Low Glucose alert to 70 mg/dL (3.9 mmol/L),
how often will your alarm alert you if your blood glucose goes below
70 mg/dL (3.9 mmol/L)? If your blood sugar goes below 70 mg/dL (3.9
mmol/L), you can expect your alarm to sound 83% of the time and not to
sound approximately 17% of time within the 15 minute period before or
after your blood sugar goes below 70 mg/dL(3.9 mmol/L).
The High Glucose Alert
Estimates of how well the adjustable High Glucose Alert performs are
presented in Table 6.
Содержание G4 Platinum Professional
Страница 13: ...12 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1...
Страница 23: ...22 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 24: ...Dexcom G4 PLATINUM Professional System User s Guide 23 2 chapter two INDICATIONS FOR USE AND SAFETY STATEMENT...
Страница 25: ...24 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 2...
Страница 35: ...34 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...Dexcom G4 PLATINUM Professional System User s Guide 35 3 chapter three RISKS AND BENEFITS...
Страница 37: ...36 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1 3...
Страница 43: ...42 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 4...
Страница 51: ...50 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 5...
Страница 65: ...64 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 6...
Страница 81: ...80 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 7...
Страница 93: ...92 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10 8...
Страница 103: ...102 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 104: ...Dexcom G4 PLATINUM Professional System User s Guide 103 9 chapter nine ALERTS ALARM PROFILES...
Страница 105: ...104 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 9...
Страница 125: ...124 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 126: ...Dexcom G4 PLATINUM Professional System User s Guide 125 10 chapter ten EVENTS DEXCOM STUDIO SOFTWARE AND BLINDED MODE...
Страница 127: ...126 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10...
Страница 140: ...Dexcom G4 PLATINUM Professional System User s Guide 139 11 chapter eleven ENDING A SENSOR SESSION...
Страница 141: ...140 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 11...
Страница 147: ...146 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 149: ...148 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 12...
Страница 164: ...Dexcom G4 PLATINUM Professional System User s Guide 163 13 chapter thirteen TROUBLESHOOTING...
Страница 165: ...164 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 13...
Страница 182: ...Dexcom G4 PLATINUM Professional System User s Guide 181 14 chapter fourteen TECHNICAL INFORMATION...
Страница 183: ...182 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 14...
Страница 218: ...Dexcom G4 PLATINUM Professional System User s Guide 217 15 chapter fifteen USER ASSISTANCE...
Страница 219: ...218 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 15...
Страница 222: ...Dexcom G4 PLATINUM Professional System User s Guide 221 16 chapter sixteen WARRANTY...
Страница 223: ...222 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 16...
Страница 231: ...230 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 232: ...Dexcom G4 PLATINUM Professional System User s Guide 231 17 chapter seventeen TRAVEL INFORMATION...
Страница 233: ...232 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 17...
Страница 235: ...234 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...
Страница 236: ...Dexcom G4 PLATINUM Professional System User s Guide 235 18 chapter eighteen APPENDIX...
Страница 237: ...236 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 18...
Страница 259: ...258 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System PAGE INTENTIONALLY LEFT BLANK...