11
SPECIFICATIONS
DEVILBISS IFILL PERSONAL OXYGEN STATION
Operating Temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41º to 95ºF (5º to 35ºC)
Operating Humidity Range: . . . . . . . . . . . . . . . . . . . . . . .15 to 95% R.H. non-condensing
Operating Altitude: . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 6,562 Feet (0 to 2,000 Meters)
Storage Temperature Range: . . . . . . . . . . . . . . . . . . .-40ºF to +158ºF (-40ºC to +70ºC)
Storage Humidity Range: . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 to 95% non-condensing
Electrical Rating: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115V~ 60Hz 4.1 Amps
Operating Voltage Range: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97 – 127V~ 60HZ
Power Range: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .400 Watts Average
Oxygen Purity: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93% O2 ± 3%
Dimensions: (including casters) w/o cylinder . . . . . . . . . . . . .12.25”W x 28.5”H x 22.5”D
Weight: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66 lbs
Shipping Weight: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76 lbs
This unit is classified as nonprotected (ordinary equipment) per EN60601-1 . . . . . .IPX0
Specifications subject to change without notice.
IFILL OXYGEN CYLINDER TYPICAL FILL TIMES
NOTE–
All filling times are approximate and may vary depending on altitude and environmental conditions.
Typical cylinder fill time from empty to 2,000 +/- 200 psig are as follows:
M4 . . . . . . . . . . . 60 Minutes
M6 . . . . . . . . . . . 75 Minutes
ML6 . . . . . . . . . .90 Minutes
C . . . . . . . . . . .130 Minutes
D . . . . . . . . . . .215 Minutes
E . . . . . . . . . . .350 Minutes
NOTE-
Degradation of performance may occur if unit is operated outside of specified operating parameters.